Calibration and cross-laboratory implementation of scintillating screens for electron bunch charge determination
Thomas Kurz1, Jurjen Pieter Couperus1, Jakob Matthias Krämer1
1Helmholtz-Zentrum Dresden-Rossendorf, Bautzner Landstraße 400, 01328 Dresden, Germany.
Abstract:
We revise the calibration of scintillating screens commonly used to detect relativistic electron beams with low average current, e.g., from laser-plasma accelerators, based on new and expanded measurements that include higher charge density and different types of screens than previous work [Buck et al., Rev. Sci. Instrum. 81, 033301 (2010)]. Electron peak charge densities up to 10 nC/mm2 were provided by focused picosecond-long electron beams delivered by the Electron Linac for beams with high Brilliance and low Emittance (ELBE) at the Helmholtz-Zentrum Dresden-Rossendorf. At low charge densities, a linear scintillation response was found, followed by the onset of saturation in the range of nC/mm2. The absolute calibration factor (photons/sr/pC) in this linear regime was measured to be almost a factor of 2 lower than that reported by Buck et al. retrospectively implying a higher charge in the charge measurements performed with the former calibration. A good agreement was found with the results provided by Glinec et al. [Rev. Sci. Instrum. 77, 103301 (2006)]. Furthermore long-term irradiation tests with an integrated dose of approximately 50 nC/mm2 indicate a significant decrease of the scintillation efficiency over time. Finally, in order to enable the transfer of the absolute calibration between laboratories, a new constant reference light source has been developed.
More Related Videos
07:31Implementation of a Real-Time Psychosis Risk Detection and Alerting System Based on Electronic Health Records using CogStack
Published on: May 15, 2020
07:52Author Spotlight: Enhancing Cryo-Electron Microscopy by Automated Data Collection and Analysis Techniques
Published on: December 1, 2023
Related Concept Videos
Formal Charges
Ions and Ionic Charges
Atomic Radii and Effective Nuclear Charge
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Electron Carriers
Over the many stages of cellular respiration, glucose breaks down into carbon dioxide and water. Electron carriers pick up electrons lost by glucose in these reactions, temporarily storing and releasing them into the electron...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...