Comparison of Coagulation Parameters in Arterial and Venous Blood in Cardiac Surgery Measured Using the Quantra
Danja S Groves1, Deborah A Winegar2, Lucas G Fernandez1
1Department of Anesthesiology, University of Virginia Health System, Charlottesville, VA.
Objective:
Perioperative coagulation testing often is performed with arterial samples even though device reference ranges typically are established in venous samples. Although limited studies exist comparing coagulation parameters across sampling sites, viscoelastic testing devices have demonstrated some differences. The objective of this study was to compare coagulation parameters determined using the Quantra System for venous and arterial samples.
Design:
Prospective, observational study.
Setting:
Tertiary care university hospital.
Participants:
The study comprised 30 adult patients undergoing cardiac surgery.
Interventions:
Paired arterial and venous samples were obtained at 2 of the following time points: baseline, during bypass, or after protamine reversal of heparin. Quantra measurements included Clot Time (CT), Heparinase Clot Time (CTH), Clot Time Ratio (CTR), Clot Stiffness (CS), and Fibrinogen (FCS) and Platelet (PCS) Contributions to clot stiffness.
Measurements And Main Results:
The relationship and agreement between matched data pairs were established and statistical analysis was performed via paired t tests. CTR, CS, FCS, and PCS were unaffected by the sampling site, whereas CT and CTH demonstrated statistically significant differences between arterial and venous samples (p < 0.001). Arterial clot times were prolonged relative to the venous ones with a mean percent error of 14.2 % and 11.9 %, respectively. These results are in general agreement with those reported for other viscoelastic testing devices.
Conclusions:
This study demonstrates that Quantra clot stiffness-based parameters (CS, FCS, PCS) are unaffected by sampling site, whereas the clot time parameters (CT and CTH) show good correlation in the presence of a bias. CTR, a ratio of CT and CTH, also is unaffected.
More Related Videos
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
08:27Adult Mouse Venous Hypertension Model: Common Carotid Artery to External Jugular Vein Anastomosis.
Published on: January 27, 2015
Related Concept Videos
Coagulation
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Measurement of Blood Pressure
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...
The Sense of Self: Reflected Self-Appraisal and Social Comparison