Incorrect Holter-ECG analysis caused by the pacemaker delivering small high-frequency currents for thoracic impedance
Masaomi Chinushi1, Hitoshi Tachikawa2, Yuko Chinushi2
1Graduate School of Health Science, Niigata University School of Medicine, Niigata, Japan.
Abstract:
In an 86-year-old woman who had been treated for sick sinus syndrome, the small high-frequency current delivered by pacemaker in order to measure the minutes ventilation for utilizing the rate-response mode was transiently over-sensed on the Holter electrocardiogram. Although her pacing system was working appropriately, the numbers of the paced beats on the automatic Holter analysis were undercounted (from >60% to <5%) during the over-sensing periods because these currents were recognized as multiple pacing spikes. Physicians need to pay attention to such multiple pacing spike markers, because these can be a cause of unreliable results of the Holter analysis. <Learning objective: Paying attention to multiple pacing spike markers on Holter electrocardiogram is clinically important in patients treated by a rate-response pacemaker. This is because (1) this does not imply the malfunction of the pacemaker system, and (2) results from automatic Holter analysis would not be reliable during the period when these multiple pacing markers were recorded.>.
Related Concept Videos
Mesh Analysis with Current Sources
Current Source in One Mesh: The analysis process is straightforward when a current source is found in only one mesh within the circuit. Mesh currents are assigned as usual, with the mesh containing the current source excluded from the analysis. Kirchhoff's voltage law...
Thoracic Aorta
Holter Monitor: 24-Hour Monitoring
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Impedance Combination
The Thoracic Cage: Sternum
The sternum is the elongated bony structure on the anterior side of the thoracic cage. It consists of three parts: the manubrium, the body, and the xiphoid...