Integrated Photoacoustic Ophthalmoscopy and Spectral-domain Optical Coherence Tomography
Published on: January 15, 2013
Measurable Range of Subfoveal Choroidal Thickness With Conventional Spectral Domain Optical Coherence Tomography
Mingui Kong1,2, Da Ye Choi3, Gyule Han3
1Hangil Eye Hospital, Incheon, Korea.
Purpose:
We investigate the measurable range of subfoveal choroidal thickness (SFCT) with conventional spectral domain optical coherence tomography (SD-OCT) in normal healthy eyes.
Methods:
All subjects underwent an SD-OCT horizontal line scan centered at the fovea with standard and enhanced depth imaging (EDI) techniques. Two independent observers manually measured SFCT on standard and EDI-OCT images, if two choroidal borders were identified. The rate of successful measurement with standard OCT was evaluated. Inter- and intraclass correlation coefficients and Bland-Altman plots were used for analysis.
Results:
OCT images of 724 normal healthy eyes from 374 subjects were evaluated. Although the inner choroidal border was identified on all OCT images, the outer choroidal border was identified on 378 of 724 (52.2%) conventional OCT images. Mean SFCT of all study eyes measured by EDI-OCT images was 292.6 ± 94.0. The number of successful measurements with conventional OCT was 100%, 87.5%, 48.1%, 33.0%, and 0%, in the SFCT ranges of <241, 241-280, 281-320, 321-360, and >360 μm, respectively. The accumulated rate of successful measurement with conventional OCT was 100%, 96.4%, and 82.2% in the SFCT ranges of ≤240, ≤280, and ≤320 μm, respectively. Two protocols showed good inter-correlation of SFCT, when SFCT were measurable in both scans.
Conclusions:
SFCT can be measured by conventional OCT in eyes with thin choroid, and the measured value shows high agreement with those measured by EDI-OCT. In such eyes, conventional OCT can be used potentially to evaluate the retina and choroid.
Translational Relevance:
In eyes with a thin choroid, conventional OCT can be used potentially to evaluate the retina and choroid with high agreement with EDI-OCT.
More Related Videos
Related Concept Videos
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Conservation of Protein Domains Over Different Proteins
A limited set of protein domains often duplicate and recombine during evolution. These domains can be organized in different combinations to...
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
Enolate Mechanism Conventions
Membrane Domains
Protein Domains
The membrane comprises a group of distinct proteins responsible for carrying out a cell's specific function. For example, the plasma membrane of the human sperm, or a single germ cell, contains a unique set of proteins in the...
Three Developmental Domains
Physical Development
Physical processes, also known as maturation, encompass the biological changes that occur across an individual's life. These changes begin with genetic inheritance and continue through various stages, including growth in height and weight,...