Impact of image reconstruction methods on quantitative accuracy and variability of FDG-PET volumetric and textural
Ali Ketabi1,2, Pardis Ghafarian3,4, Mohammad Amin Mosleh-Shirazi5
1Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
Objective:
This study aims to assess the impact of different image reconstruction methods on PET/CT quantitative volumetric and textural parameters and the inter-reconstruction variability of these measurements.
Methods:
A total of 25 oncology patients with 65 lesions (between 2017 and 2018) and a phantom with signal-to-background ratios (SBR) of 2 and 4 were included. All images were retrospectively reconstructed using OSEM, PSF only, TOF only, and TOFPSF with 3-, 5-, and 6.4-mm Gaussian filters. The metabolic tumor volume (MTV) and total lesion glycolysis (TLG) were measured. The relative percent error (ΔMTV and ΔTLG) with respect to true values, volume recovery coefficients, and Dice similarity coefficient, as well as inter-reconstruction variabilities were quantified and assessed. In clinical scans, textural features (coefficient of variation, skewness, and kurtosis) were determined.
Results:
Among reconstruction methods, mean ΔMTV differed by -163.5 ± 14.1% to 6.3 ± 6.2% at SBR2 and -42.7 ± 36.7% to 8.6 ± 3.1 at SBR4. Dice similarity coefficient significantly increased by increasing SBR from 2 to 4, ranging from 25.7 to 83.4% between reconstruction methods. Mean ΔTLG was -12.0 ± 1.7 for diameters > 17 mm and -17.8 ± 7.8 for diameters ≤ 17 mm at SBR4. It was -31.7 ± 4.3 for diameters > 17 mm and -14.2 ± 5.8 for diameters ≤ 17 mm at SBR2. Textural features were prone to variations by reconstruction methods (p < 0.05).
Conclusions:
Inter-reconstruction variability was significantly affected by the target size, SBR, and cut-off threshold value. In small tumors, inter-reconstruction variability was noteworthy, and quantitative parameters were strongly affected. TOFPSF reconstruction with small filter size produced greater improvements in performance and accuracy in quantitative PET/CT imaging.
Key Points:
• Quantitative volumetric PET evaluation is critical for the analysis of tumors. • However, volumetric and textural evaluation is prone to important variations according to different image reconstruction settings. • TOFPSF reconstruction with small filter size improves quantitative analysis.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Shape and Texture of Coarse Aggregate
Improving Translational Accuracy
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Structures of Solids
Network Covalent Solids
To break or to melt a covalent network solid, covalent bonds must be broken. Because covalent bonds are relatively strong, covalent network solids are typically...


![Visualization and Quantification of Brown and Beige Adipose Tissues in Mice using [18F]FDG Micro-PET/MR Imaging](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F62460.jpg&w=3840&q=50)