Reducing radiation exposure by lowering frame rate in children undergoing cardiac catheterization: A quality improvement study
Related Experiment Video
Updated: Feb 4, 2026

Studying Chronic Exposure of Mice to Ultraviolet B Radiation
Published on: August 19, 2025
Shahnawaz M Amdani1, Robert D Ross1, Paul A Webster1
1Division of Cardiology, Carman and Ann Adams Department of Pediatrics, Children's Hospital, Wayne State University School of Medicine, Detroit, Michigan.
Insights
Lowering frame rates in pediatric cardiac catheterization significantly reduced radiation dose without increasing procedure time, contrast volume, or complications. Diagnostic image quality was maintained, ensuring safe and effective procedures.
Area of Science:
- Pediatric Cardiology
- Medical Imaging
- Radiation Safety
Background:
- Radiation dose reduction is critical in pediatric cardiac catheterization to minimize stochastic effects in children with congenital heart disease.
- Lowering frame rates may reduce radiation but could compromise image quality, potentially increasing fluoroscopy time and complications.
Purpose of the Study:
- To assess the impact of reduced frame and pulse rates on radiation dose, fluoroscopic time, contrast volume, image quality, and complication rates in the pediatric cardiac catheterization laboratory.
- To determine if lowering frame rates is a viable strategy for radiation dose reduction without compromising procedural outcomes.
Main Methods:
- Retrospective analysis of data from three eras with varying cineangiography (frame/sec) and fluoroscopy (pulse/sec) rates.
- Evaluation of image quality by blinded operators.
- Assessment of radiation dose, fluoroscopy time, contrast volume, and complication rates.
Main Results:
- Overall radiation dose significantly decreased across the eras (70.0 to 36.6 µGym²/kg).
- No significant increase in fluoroscopy time, contrast volume, or diagnostic image quality was observed between 30 and 15 frames/sec.
- No difference in significant adverse event rates was found.
Conclusions:
- Lowering frame and pulse rates in pediatric cardiac catheterization is an effective quality improvement strategy.
- This approach successfully reduced radiation dosage without compromising procedural efficacy, safety, or diagnostic image quality.
Introduction:
Reduction of radiation dosage in the pediatric cardiac catheterization laboratory (PCL) is important to reduce the risk of its stochastic effect in children with congenital heart disease. Lowering the frame rate would reduce radiation dosage possibly at the expense of image quality, potentially resulting in higher fluoroscopic time and procedural complication rate.
Methods:
The data were retrospectively analyzed in three eras: era 1 (n = 234), cineangiography 30 frames/sec (f/s) and fluoroscopy 15 pulse/sec (p/s); era 2 (n = 381), cineangiography 30 f/s and fluoroscopy 6 p/s; and era 3 (n = 328), cineangiography 15 f/s and fluoroscopy 6 p/s. Also, three operators blinded to the frame rate setting evaluated cineangiography image quality. In this study, the impact of lowering the default frame rates on radiation dosage, fluoroscopic time, contrast volume, diagnostic image quality, and complication rates in the PCL was assessed.
Results:
Overall radiation dosage progressively declined during these eras (70.0 vs 64.1 vs 36.6 µGym2 /kg, P < .001) without a difference in significant adverse event rates. There was no significant increase in either fluoroscopy time or contrast volume. There was no difference in the diagnostic image quality between cineangiography 30 and 15 f/s. Lowering the default frame/pulse rates of both fluoroscopy and cineangiography significantly decreased the overall radiation dosage in the PCL. Importantly, fluoroscopy time, contrast volume, and complication rates did not increase while maintaining diagnostic image quality.
Conclusion:
This quality improvement project proved successful in lowering radiation dosage without compromising the efficacy and safety of catheterizations.