[Application of Two-Dimensional Near-Infrared Correlation Spectroscopy in the Specificity Analysis of Noninvasive
Abstract:
In this paper, two-dimensional (2D) correlation spectroscopy analysis was applied to investigate the influence of the main component in blood and the systematic drift during the measurement on the specificity of glucose in the near-infrared (NIR) spectroscopy. First, the NIR transmittance of glucose aqueous solutions was measured and the 2D correlation NIR spectra were calculated under the perturbation of glucose concentration. Based on the comparative analysis for synchronous and asynchronous 2D correlation spectra, the characteristic absorption peaks of glucose in the combination band and the overtone band were determined. Then a small amount of albumin was added into glucose aqueous solutions, and the transmittance was recorded to perform 2D correlation spectroscopy analysis under the perturbation of glucose concentration. However, the absorption of glucose in the first overtone band (1590nm) and second overtone band (1195nm) was no longer homologous in the 2D correlation spectra, which means that the albumin may reduce the specificity of glucose. Further, the oral glucose tolerance test of healthy volunteer was conducted and the NIR diffuse reflectance of left palm was collected in vivo. The 2D correlation analysis results showed that, the homology of glucose in the diffuse reflectance was also destroyed. Moreover, as the spectral variation from the glucose concentration change is too low to be covered by that induced by systematic drift easily, some background correction methods were usually required. For the transmittance experiment of glucose aqueous solutions and the diffuse reflectance experiment of human body, the pure water sample and 5% diffuse reflectance standard were used as the reference, respectively. Then 2D correlation spectroscopy was developed under the perturbation of measurement time. Results showed that, smaller band shift was observed in the slice spectra of 2D correlation synchronous spectra after the corresponding background correction, and the specificity of glucose was improved both in the in vitro and in vivo experiments. So for the non-invasive glucose sensing by NIR spectroscopy, the wavelengths should be chosen carefully to avoid the absorption band of some interfering components which may destroy the homology of glucose and make spectral interpretation more complicated. And the selection of reference samples for relative measurement is also important to improve the specificity of glucose.
More Related Videos
10:50Detection and Quantification of Plasmodium falciparum in Aqueous Red Blood Cells by Attenuated Total Reflection Infrared Spectroscopy and Multivariate Data Analysis
Published on: November 2, 2018
07:27How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Related Concept Videos
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Dimensional Analysis
In fluid mechanics, dimensional...
Dimensional Analysis
Dimensional analysis allows us to analyze and compare physical quantities on a...
Dimensional Analysis