Related Experiment Video
Updated: Feb 4, 2026

Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
Long-Term Outcome of Hearing Rehabilitation With An Active Middle Ear Implant
Faris F Brkic1, Dominik Riss1, Alice Auinger1
1Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Vienna, Vienna, Austria.
Objective:
To assess the audiological and long-term medical and technical follow-up outcomes of an active middle ear implant.
Methods:
This was a retrospective medical chart analysis of all patients provided with an active middle ear implant in a tertiary academic medical referral center between September 1, 1998, and July 31, 2015. Main outcome measures were medical and technical complications, revisions, reimplantations, explantations, coupling approaches, mean time of use, pre- and postoperative hearing thresholds, functional hearing gain across frequencies (250-4,000 Hz), and Freiburg monosyllablic word test at 65 dB.
Results:
One hundred and three patients were identified. Fifteen were implanted bilaterally (n = 118 Vibrant Soundbridge devices [MED-EL, Innsbruck, Austria]). Seventy-seven devices were implanted for sensorineural and 41 for mixed and conductive hearing loss. Patients used the implant for 6.7 years (range 0.7 months-17.9 years) on average. Ninety-one patients (77.12%) were using the device at the end of the observation period. An overall complication rate of 16.1% was observed. The revision and explantation rates were higher for devices implanted between 2004 and 2006. The device failure rate was 3.4%. Audiological evaluation showed significant hearing gains for both hearing loss patient groups.
Conclusion:
This long-term follow-up reveals the reliability of the active middle ear implant in a single center. Overall complication rate and device failure rate are acceptable. The complication rate was higher during implementation of alternative coupling approaches. The audiological benefit was satisfactory in patients with all hearing loss types. The majority of implanted patients used the implant at the end of the observation period.
Level Of Evidence:
4 Laryngoscope, 129:477-481, 2019.
Related Concept Videos
Hearing
Anatomy of the Ear
Predicting Reaction Outcomes
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...
Long-term Depression
Guidelines for Writing Outcome
Patient outcomes reflect the patient's response to the goal rather than what the nurse aims to achieve. Terminology should be observable and measurable to avoid the reader's interpretation. The desired outcome should be realistic and achievable in the designated care timeframe. Expected outcomes should align with adjunctive therapies. The outcome should enhance care...