Effect of microscopic susceptibility gradients on chemical-shift-based fat fraction quantification in supraclavicular
Drew McCallister1,2, Le Zhang3,2, Alex Burant1,2
1Department of Physics and Astronomy, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA.
Background:
Susceptibility differences between fat and water can cause changes in the water-fat frequency separation that can negatively affect the accuracy of fat fraction techniques. This may be especially relevant for brown adipose tissue, as MRI fat fraction techniques have been proposed for its detection.
Purpose:
To assess the effect of microscopic magnetic susceptibility gradients on the water-fat frequency separation and its impact on chemical-shift-based fat fraction quantification techniques in the supraclavicular fat, where brown adipose tissue is commonly found in humans.
Study Type:
Prospective.
Population/Subjects/Phantom/Specimen/Animal Model:
Subjects: 11 healthy volunteers, mean age of 26 and mean BMI of 23, three overweight volunteers, mean age of 38 and mean BMI of 33. Phantoms: bovine phantom and intralipid fat emulsion. Simulations: various water-fat distributions.
Field Strength/Sequence:
Six-echo gradient echo chemical-shift-encoded sequence at 3T.
Assessment:
Fat fraction values as obtained from a water-fat spectral model accounting for susceptibility-induced water-fat frequency variations were directly compared to traditional spectral models that assume constant water-fat frequency separation.
Statistical Tests:
Two-tail t-tests were used for significance testing (p < 0.05.) A Bayesian Information Criterion difference of 6 between fits was taken as strong evidence of an improved model.
Results:
Phantom experiments and simulation results showed variations of the water-fat frequency separation up to 0.4 ppm and 0.6 ppm, respectively. In the supraclavicular area, the water-fat frequency separation produced by magnetic susceptibility gradients varied by as much as ±0.4 ppm, with a mean of 0.08 ± 0.14 ppm, producing a mean difference in fat fraction of -1.26 ± 5.26%.
Data Conclusion:
In the supraclavicular fat depot, microscopic susceptibility gradients that exist within a voxel between water and fat compartments can produce variations in the water-fat frequency separation. These variations may produce fat fraction quantification errors of 5% when a spectral model with a fixed water-fat frequency separation is applied, which could impact MR brown fat techniques.
Level Of Evidence:
1 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;49:141-151.
Related Concept Videos
Fats as Energy Storage Molecules
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
Inductive Effects on Chemical Shift: Overview
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
π Electron Effects on Chemical Shift: Overview