Performance evaluation of five commercial assays for detection of acetaminophen
Bao-Yum Chan1, Hing-Man Tsang1, Candy Wai-Yan Ng2
1Department of Pathology, Clinical Laboratory, North Lantau Hospital, Lantau, Hong Kong.
Background:
To evaluate the analytical performance of five commercial acetaminophen assays and select the best method for routine use.
Methods:
Imprecision, accuracy, linearity, and interferences of three enzymatic assays (Beckman Coulter AU Paracetamol, Abbott MULTIGENT Acetaminophen, and Sekisui Acetaminophen L3K) and two immunoassay-based assays (Beckman Coulter SYNCHRON ACTM (Acetaminophen) Reagent and Siemens SYVA Emit-tox Acetaminophen) were evaluated on a Beckman Coulter AU680 chemistry analyzer. Hook effect for immunoassay-based assays and recovery in ultrafiltrate for enzymatic methods were studied.
Results:
Within-run and between-run imprecision of the enzymatic assays ranged 0.26%-0.82% and 0.53%-2.86%, respectively, while that for the immunoassay-based methods ranged 0.96%-6.34% and 1.50%-11.33%, respectively. All assays except the SYNCHRON assay fell within the program analytical performance specifications (±20 µmol/L or 10%) for external quality assurance (EQA) samples, with the highest positive bias (31.7%) observed in the SYNCHRON assay. Icteric interference was demonstrated most significantly in the Abbott assay (up to 88 μmol/L positive bias in blank serum). The lipemic interference on the SYNCHRON was significant (up to 110% positive bias at level of 100 μmol/L). The immunoassay-based methods were less susceptible to hemolytic interference, while the Abbott and AU assays were more susceptible to N-acetylcysteine interference. Both immunoassay-based methods showed no hook effect up to 18 000 μmol/L. Ultrafiltration recoveries for enzymatic methods were satisfactory, ranging from 80.0% ± 5.1% to 89.5% ± 3.0%.
Conclusions:
Proportional bias was observed in the SYNCHRON assay, while the Siemens and Sekisui assays were minimally affected by bilirubin interferences.
More Related Videos
09:44Generation of a Rat Model of Acute Liver Failure by Combining 70% Partial Hepatectomy and Acetaminophen
Published on: November 27, 2019
08:26A Comprehensive Procedure to Evaluate the In Vitro Performance of the Putative Hemangioblastoma Neovascularization Using the Spheroid Sprouting Assay
Published on: April 12, 2018
Related Concept Videos
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...
Pathophysiology of Cardiac Performance
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
Sources of Self-Esteem II: Performance Feedback
Self-Evaluation: Self-Enhancement and Self-Verification