1Division of Imaging, Diagnostics, and Software Reliability, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, 10903 New Hampshire Avenue, Building 62, RM 4126, Silver Spring, MD.
Rationale And Objectives:
The quantitative assessment of volumetric CT for discriminating small changes in nodule size has been under-examined. This phantom study examined the effect of imaging protocol, nodule size, and measurement method on volume-based change discrimination across low and high object to background contrast tasks.
Materials And Methods:
Eight spherical objects ranging in diameter from 5.0 mm to 5.75 mm and 8.0 mm to 8.75 mm with 0.25 mm increments were scanned within an anthropomorphic phantom with either foam-background (high-contrast task, ∼1000 HU object to background difference)) or gelatin-background (low-contrast task, ∼50 to 100 HU difference). Ten repeat acquisitions were collected for each protocol with varying exposures, reconstructed slice thicknesses and reconstruction kernels. Volume measurements were obtained using a matched-filter approach (MF) and a publicly available 3D segmentation-based tool (SB). Discrimination of nodule sizes was assessed using the area under the ROC curve (AUC).
Results:
Using a low-dose (1.3 mGy), thin-slice (≤1.5 mm) protocol, changes of 0.25 mm in diameter were detected with AU = 1.0 for all baseline sizes for the high-contrast task regardless of measurement method. For the more challenging low-contrast task and same protocol, MF detected changes of 0.25 mm from baseline sizes ≥5.25 mm and volume changes ≥9.4% with AUC≥0.81 whereas corresponding results for SB were poor (AUC within 0.49-0.60). Performance for SB was improved, but still inconsistent, when exposure was increased to 4.4 mGy.
Conclusion:
The reliable discrimination of small changes in pulmonary nodule size with low-dose, thin-slice CT protocols suitable for lung cancer screening was dependent on the inter-related effects of nodule to background contrast and measurement method.
More Related Videos
Related Concept Videos
Work Done During Volume Change
Consider a gas confined to a cylinder fitted with a movable piston at one end. If the gas expands from volume V1 to volume V2, it exerts a force on the piston, such that the piston moves by a distance dr.
The work done by the gas on the piston can be expressed as
Unsoundness of Aggregate due to Volume Change
Le Chatelier's Principle: Changing Volume (Pressure)
Stereotypes, Prejudice, and Discrimination
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT