Is capillary zone electrophoresis a suitable method for estimating serum albumin: A comparison of four methods
Maël Padelli1, Cyril Leven1, Eymeric Le Reun2
1Department of Biochemistry and Pharmaco-Toxicology, University Hospital of Brest, Brest, France; Univ Brest, LIEN, F-29200 Brest, France.
Background:
The capillary zone electrophoresis method of albumin measurement is frequently used for oncologic and haematologic patients but few data exist about the agreement between the albumin measurements performed by capillary zone electrophoresis and other methods. The aim of this study was to analyse the agreement between human serum albumin measurements by capillary zone electrophoresis and by the nephelometry, bromocresol purple and turbidimetry methods.
Method:
We prospectively measured 100 freshly collected non-frozen patient serum samples, by using four different methods: the capillary zone electrophoresis method performed with a CAPILLARYS 2 instrument, the bromocresol purple dye method performed with an Advia XPT analyser, the nephelometric method performed with a BN ProSpec analyser and the turbidimetric method with reagents from DiAgam and performed with the Advia XPT analyser.
Results:
A bias towards higher values in the lower range of albumin concentrations was observed with capillary zone electrophoresis compared to immunonephelometry: correlation coefficient r2 = 0.925; slope of 0.86 (0.82-0.89, 95% confidence interval), which is significantly different from 1; and an intercept of 4.94 g/L (3.67-6.16, 95% confidence interval). Similar results were observed when comparing capillary zone electrophoresis to the bromocresol purple and immunoturbidimetry methods. The capillary electrophoresis method overestimated low albumin levels by up to 25% (5 g/L).
Conclusion:
Compared to the nephelometry, turbidimetry and bromocresol purple methods, the capillary zone electrophoresis method tends to overestimate human serum albumin concentrations for levels below 30 g/L. This discrepancy could lead to an overestimation of the nutritional status, an inappropriate scoring of the disease and a delay in nutritional treatment.
More Related Videos
14:53Capillary Electrophoresis Separation of Monoclonal Antibody Isoforms Using a Neutral Capillary
Published on: January 16, 2017
07:54Analysis of AtHIRD11 Intrinsic Disorder and Binding Towards Metal Ions by Capillary Gel Electrophoresis and Affinity Capillary Electrophoresis
Published on: August 22, 2018
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Hybrid Zones
One-Compartment Open Model: Wagner-Nelson and Loo Riegelman Method for ka Estimation
On...
Zones of Protection
Protective zones are defined by closed dashed lines, containing one or more components. A key characteristic of these zones is the strategic placement of...