Effect of electrode separation on high-resolution impedance manometry catheters for ex vivo animal experiments
Albert Ruiz-Vargas1, Reizal Mohd Rosli2, Lukasz Wiklendt3
1The Medical Device Research Institute, College of Science and Engineering, Flinders University, Adelaide, Australia.
Background:
High-resolution impedance manometry (HRIM) catheters are used for esophageal reflux and bolus transit studies. Usually, these catheters have electrodes spaced at 20 mm intervals that are used for measuring electrical impedance between the consecutive electrodes and pressure sensors with 10 mm separation. Electrical impedance is a distributed measurement between the electrodes, unlike the point measurement of pressure sensors. Thus, the electrode separations affect the overall spatiotemporal resolution of the measurement.
Methods:
A 3D model of the gastrointestinal tract, in which bolus shape and size can be modified, was used to simulate the admittance signal response of an intraluminal impedance catheter with 10 and 20 mm electrode separation to study the distributed sensitivity in the lumen to gain insight on the impedance measurement. In addition, experiments on sections of rabbit proximal colon were conducted with two catheters with electrode separations of 10 and 20 mm to compare the experimental data with the simulated data.
Key Results:
Reducing electrode spacing from 20 to 10 mm increased sensitivity to diameter change by a factor of ten. Admittance and diameter correlated strongly during a myogenic contraction with a Pearson's Correlation Coefficient of 0.86 for the custom catheter, in comparison with 0.56 for the commercial HRIM catheter.
Conclusions:
Ten millimeter electrode separation has a better spatiotemporal resolution, and unlike 20 mm electrode separation is able to identify myogenic contractions. Based on the numerical and experimental data, closer electrode separation should be considered for improved spatial resolution.
Related Concept Videos
Impedance Combination
Standard Electrode Potentials
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
What is an Experiment?
Series Impedances: Three-Phase Line
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...
Bus Impedance Matrix
In the first circuit, all machine voltage sources are short-circuited, leaving only the prefault voltage source at the fault location. The positive-sequence bus impedance matrix can be determined by solving the nodal equations,...