[Possibilities for residual hearing preservation with Nucleus CI532 Slim Modiolar electrode array. Case report]
Roland Nagy1, János András Jarabin1, Balázs Dimák1
1Általános Orvostudományi Kar, Fül-Orr-Gégészeti és Fej-Nyaksebészeti Klinika, Szegedi Tudományegyetem Szeged, Tisza Lajos krt. 111., 6725.
Orvosi Hetilap
|October 9, 2018
Summary
This study shows that the Cochlear® Nucleus CI532 Slim Modiolar electrode can preserve residual hearing after cochlear implantation. Minimally invasive surgery with this electrode helps maintain hearing function for improved patient outcomes.
Area of Science:
- Otolaryngology
- Neurosurgery
- Audiology
Background:
- Preserving residual acoustic hearing during cochlear implantation (CI) is crucial for enhancing hearing performance in hearing-impaired patients.
- Minimally invasive surgical techniques and atraumatic electrode designs are essential for minimizing implantation trauma.
Observation:
- A case study involving a patient with congenital hearing loss and residual hearing who received a Cochlear® Nucleus CI532 Slim Modiolar electrode.
- The electrode was implanted via a minimally invasive round window approach on the right ear.
- Audiological monitoring was conducted to assess postoperative acoustic residual hearing.
Findings:
- Postoperative hearing thresholds showed an initial progression of 20-25 dBHL at lower frequencies (0.25-1.0 kHz) and 5-10 dBHL at higher frequencies (2.0-4.0 kHz) compared to preoperative levels.
- Hearing thresholds at the 6-month mark indicated a slight progression above 1 kHz, but by 12 months, they improved to levels comparable to the 4th postoperative week.
- The slim modiolar electrode's proximity to the modiolus may reduce hydrodynamic load and offer neuroprotection by enabling lower stimulation intensity.
Implications:
- Preservation of acoustic residual hearing post-CI can significantly improve speech perception and sound localization abilities.
- Maintaining residual hearing is vital for future regenerative procedures and potential drug treatments for hearing loss.
- The findings support the use of slim modiolar electrodes in cochlear implantation for patients with residual hearing.
Related Concept Videos
Hearing
57.2K
When we hear a sound, our nervous system is detecting sound waves—pressure waves of mechanical energy traveling through a medium. The frequency of the wave is perceived as pitch, while the amplitude is perceived as loudness.
57.2K
The Nucleus
102.4K
The nucleus is a membrane-bound organelle that acts as a control center in a eukaryotic cell. It contains chromosomal DNA, which controls gene expression and precisely regulates the production of proteins within the cell. In contrast, the DNA inside the mitochondria and chloroplast only carries out functions that are specific to those organelles.
Arrangement of DNA within Nucleus
The regulation of gene expression inside the nucleus is dependent on many factors, including the DNA structure. The...
Arrangement of DNA within Nucleus
The regulation of gene expression inside the nucleus is dependent on many factors, including the DNA structure. The...
102.4K
The Nucleus
7.6K
The nucleus is a membrane-bound organelle that acts as a control center in a eukaryotic cell. It contains chromosomal DNA, which controls gene expression and precisely regulates the production of proteins within the cell. In contrast, the DNA inside the mitochondria and chloroplast only carries out functions that are specific to those organelles.
Arrangement of DNA within Nucleus
The regulation of gene expression inside the nucleus is dependent on many factors, including the DNA structure. The...
Arrangement of DNA within Nucleus
The regulation of gene expression inside the nucleus is dependent on many factors, including the DNA structure. The...
7.6K
Residual Plots
6.5K
A residual plot is a statistical representation of data used to analyze correlation and regression results. It helps verify the requirements for drawing specific conclusions about correlation and regression. To obtain the residual plot, first, the residual for each data value is calculated, which is simply the vertical distance between the observed and the predicted value obtained from the regression equation.
When the residual values are plotted against the variable x, it is called a residual...
When the residual values are plotted against the variable x, it is called a residual...
6.5K
Residual Stresses
640
Residual stresses reside in a structure even after removing the original stress inducer. This phenomenon often arises from varied plastic deformations across different parts of a structure. Consider a rod stretched beyond its yield point. It will not regain its original length due to permanent deformation. Even after load removal, the rod does not entirely lose stress because of uneven plastic deformations, resulting in residual stresses. The computation of these stresses in structures is...
640
Standard Electrode Potentials
50.3K
On comparing the reactivity of silver and lead, it is observed that the two ionic species, Ag+ (aq) and Pb2+ (aq), show a difference in their redox reactivity towards copper: the silver ion undergoes spontaneous reduction, while the lead ion does not. This relative redox activity can be easily quantified in electrochemical cells by a property called cell potential. This property is commonly known as cell voltage in electrochemistry, and it is a measure of the energy which accompanies the charge...
50.3K