Comparison of two biometers: A swept-source optical coherence tomography and an optical low-coherence reflectometry
Hussam El Chehab1, Emilie Agard1,2, Corinne Dot1,2
1Department of Ophthalmology, Desgenettes Military Hospital, Lyon, France.
Purpose:
The technological improvements in optical biometers have made cataract surgery evolve to the level of refractive surgery. This study assessed the correlation and agreement between a biometer using swept-source optical coherence tomography technology (IOLMaster 700; Carl Zeiss Meditec, Jena, Germany) and a biometer using optical low-coherence reflectometry (Lenstar; Haag Streit, Koeniz, Switzerland).
Procedures:
Optical biometric measurements were taken with the IOLMaster 700 and the Lenstar on 129 eyes of 129 patients referred for cataract surgery. We compared biometric data (axial length, mean keratometry, central pachymetry, anterior chamber depth, lens thickness, and horizontal white-to-white measurement) and the emmetropic intraocular lens power calculated with SRK/T and Haigis formulas. Correlation and agreement for these parameters were evaluated.
Results:
Only anterior chamber depth measurements (3.05 ± 0.07 vs 3.07 ± 0.07, p < 0.001) and the white-to-white measurement (11.97 ± 0.07 vs 12.06 ± 0.07, p < 0.001) differed significantly between the machines. Correlations and agreements were excellent, with perfect agreement for axial length measurement (r = 1, intraclass correlation coefficient = 1). Emmetropic intraocular lens power did not differ between devices with SRK/T formula (20.94 ± 0.51 vs 20.92 ± 0.50, p = 0.51). With Haigis formula, there was a statistically significant difference (21.04 ± 0.52 vs 20.84 ± 0.52, p < 0.001).
Conclusion:
This study suggests that there is a very high correlation and agreement for the biometric measurements and calculation of intraocular lens implant power with the SRK/T formula.
Related Concept Videos
Sources of Self-Esteem III: Social Comparison
Properties of Enantiomers and Optical Activity
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons