[Positioning of external pelvic stabilization devices in patients with multiple injuries : Retrospective computed
G Jansen1,2, T Hefke3, G Wittenberg3
1Klinik für Anästhesiologie, Intensiv‑, Notfall‑, Transfusionsmedizin und Schmerztherapie, Evangelisches Klinikum Bethel, Burgsteig 13, 33617, Bielefeld, Deutschland. gerritjansen@web.de.
Background:
Severe hemorrhage is a dreaded complication of pelvic fractures. It has a significant impact on early trauma-associated mortality. Hemorrhage that is secondary to pelvic fractures can be reduced by external stabilization devices. Despite the commercial availability of many different systems, they are infrequently used. The aim of this computed tomography (CT) study was to examine the use of external pelvic stabilization devices.
Methods:
Between 1 January 2011 and 31 December 2015 a total of 982 images produced in CT trauma scans at a level 1 trauma centre were retrospectively examined with respect to the presence of external pelvic stabilizers. The type of device applied, its actual position including deviation from optimal position as well as pelvic parameters and complications were determined.
Results:
In 67 out of 982 patients (6.82%) with suspected multiple trauma, an external pelvic stabilizer was employed. In 41.8% the devices were not placed in concordance with prevailing scientific knowledge, 53.73% of devices did not comply with the manufacturer's instructions and 51.85% of systems with pneumatic cuffs caused significant malrotation. In one patient the cuff induced hypoperfusion of the leg but without further sequelae.
Conclusion:
While the prehospital use of pelvic slings is increasing, misplacement is very common. Especially inconsistencies between manufacturers' manuals and current scientific knowledge warrant further improvement. In systems with pneumatic cuffs malrotation of the device is common and clinically relevant. Hypoperfusion of the lower extremities is possible and should be taken into account when employing these devices.
Related Concept Videos
RNA Stability
Design Example: Identifying the Locations of Monuments in the Field Using Global Positioning System Device
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
Nuclear Stability
To hold positively charged protons together...
Muscles of the Pelvic Floor and Perineum
Perineal Layer
The perineum is a diamond-shaped area below the pelvic diaphragm, divided into an anterior urogenital triangle that contains the external genitals and a posterior anal triangle housing the anus. The urogenital...
Stability
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...