Agreement between left ventricular ejection fraction assessed in patients with gated IQ-SPECT and conventional
Lucia Leva1, Roberta Matheoud2, Gianmauro Sacchetti1
1Department of Nuclear Medicine, University Hospital 'Maggiore della Carità', Novara, Italy.
Background:
The aim of the study was to assess the agreement between the left ventricular ejection fraction (LVEF) values obtained with IQ-SPECT and those obtained with a conventional gamma camera equipped with low-energy high-resolution (LEHR), considered as the method of reference.
Methods:
Gated-stress MPI using 99mTc-tetrofosmin was performed in 55 consecutive patients. The patients underwent two sequential acquisitions (Method A and B) performed on Symbia-IQ SPECT with different acquisition times and one (Method C) on a Ecam SPECT equipped with LEHR collimators. The values of the different datasets were compared using the Bland-Altman analysis method: the bias and the limits of agreement (LA) were estimated in a head-to-head comparison of the three protocols.
Results:
In the (Method A-Method C) comparison for LVEF, the bias was 3.8% and the LAs ranged from - 9.3% to 16.8%. The agreement was still lower between Method B and C, whilst only slightly improved when Methods A and B were compared.
Conclusions:
The wide amplitude in LA intervals of about 30% indicates that IQ and LEHR GSPECT are not interchangeable. The values obtained with IQ-SPECT should only be used with caution when evaluating the functional state of the heart.
More Related Videos
09:05Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
Underflow Gates
Ligand-Gated Ion Channel Receptor: Gating Mechanism
Enolate Mechanism Conventions
Non-gated Ion Channels
Compared to the gated ion channels, the non-gated channels, also known as leakage or passive channels, have no gating mechanism....
Subcellular Fractionation
Differential Centrifugation
Differential centrifugation is...