Comparison of Metal Artifact Reduction in Dual- and Single-Source CT: A Vertebral Phantom Study
Philippe Jagoda1, Daniel Schmitz1, Stefan Wagenpfeil2
11 Saarland University Medical Center, Clinic for Diagnostic and Interventional Radiology, Kirrberger St 100, 66421 Homburg/Saar, Germany.
Objective:
The objective of this study was to compare the capability of two algorithms for metal artifact reduction and virtual monoenergetic imaging (VME), a metal artifact reduction application for dual-source CT.
Materials And Methods:
A bovine vertebra phantom with 16 artificial osteolyses and two 20 × 4.5 mm stainless steel screws was scanned on two single-source CT scanner and one dual-source CT scanner at a dose identical to the single-source acquisitions. Datasets were reconstructed with a metal artifact reduction algorithm for orthopedic implants (O-MAR, Philips Healthcare), an iterative metal artifact reduction algorithm (iMAR, Siemens Healthineers), and VME. Blinded to the method used for artifact reduction, three independent observers evaluated datasets regarding the extent of metal artifacts using a 4-point scale. Depicted osteolyses were counted and screw diameters measured for each reconstruction. Interobserver variability was evaluated using the Kendall coefficient of concordance for ordinal variables and the intraclass correlation coefficient for continuous data.
Results:
VME showed the best metal artifact reduction capability among evaluated methods; overall artifacts were rated 1.08 ± 0.29 for VME, 3.33 ± 0.65 for iMAR, and 3.91 ± 0.29 for O-MAR (p < 0.01). VME resulted in better representation of the cortical bone, trabecular structure, and soft tissue compared with the other two algorithms. VME provided the most realistic reconstruction of screw diameter. However, VME missed osteolyses. Good to almost perfect agreement was achieved for nearly all evaluated attributes.
Conclusion:
In our vertebral phantom, VME led to the most detailed representation of the osteosynthesis screw, caused the lowest amount of artifact, and represented the adjacent tissue best. Thus, VME should be considered as an alternative method to evaluate implants when other algorithms fail.
More Related Videos
10:57Synthesis and Performance Characterizations of Transition Metal Single Atom Catalyst for Electrochemical CO2 Reduction
Published on: April 10, 2018
08:59Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
Related Concept Videos
Sources of Self-Esteem III: Social Comparison
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Oxidation-Reduction Reactions
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...