A Prospective Evaluation of the Repeatability and Reliability of New Steady-state Pattern Electroretinogram
Kevin Gillmann1, Kaweh Mansouri1,2, Harsha L Rao3
1Glaucoma Research Center, Montchoisi Clinic, Swiss Vision Network, Lausanne, Switzerland.
Purpose:
The purpose of this study was to determine the repeatability and reproducibility of a new Pattern Electroretinography (PERG) device in healthy subjects.
Methods:
A total of 40 eyes of 20 healthy subjects (45.6±6.8 y, 70% female) were enrolled in this prospective study. Steady-state PERG were recorded using the Diopsys NOVA (Diopsys Inc., NJ) to obtain the Magnitude, MagnitudeD, MagnitudeD/Magnitude Ratio, and Signal to Noise Ratio. Subjects underwent 2 "contrast sensitivity" protocols with the same electrodes and 2 after changing electrodes, by a single examiner. Eyes were stimulated separately using high and low contrast patterns. The same examination was repeated by a different operator following the same protocol 6 months after the baseline measures. Repeatability and reproducibility of the readings were calculated using intraclass correlation coefficient (ICC). A Bland-Altman plot was used to analyze the agreement between measures.
Results:
ICC of Magnitude, MagnitudeD, MagnitudeD/Magnitude ratio, and SNR were 0.87, 0.81, 0.36, and 0.81, respectively between measures taken with the same electrodes (repeatability). After changing electrodes, ICC of the Magnitude, MagD, MagD/Mag ratio, and SNR were 0.79, 0.85, 0.51, and 0.83, respectively during the same session as baseline measures, and 0.87, 0.85, 0.40, and 0.76 when repeated after 6 months (reproducibility). The variability of MagD/Mag ratio increased with successive test repetitions.
Conclusions:
Mag, MagD, and SNR were repeatable and reproducible, whereas the MagD/Mag ratio was less repeatable and reproducible. These results suggest that the former parameters are sufficiently reliable to be used in clinical practice.
More Related Videos
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Reliability and Validity
Steady State Concentration
Most drugs are administered in repeated doses at fixed intervals or through continuous...
Steady Flow of a Fluid Stream
During this process, the momentum of the fluid within the control volume remains constant over the time interval dt. By applying the...
Transient and Steady-state Response
These test signals are integral in designing control systems to exhibit two key performance aspects: transient response and steady-state...
Velocity and Acceleration in Steady and Unsteady Flow
The acceleration can be generalized to any point in the flow, and expressed as components along three perpendicular directions, representing changes in velocity over...