High Efficiency Removal of Cytokines and HMGB-1 by Continuous Hemofiltration With a Dual Layered Polyethersulfone Membrane: An Ex Vivo Study
Related Experiment Video
Updated: Feb 4, 2026

Author Spotlight: Extraction of Guinea Pig Round Window Membrane to Facilitate Inner Ear Drug Delivery Research
Published on: February 23, 2024
Yoshifumi Ohchi1, Koji Goto1, Norihisa Yasuda1
1Department of Anesthesiology and Intensive Care, Oita University Faculty Of Medicine Graduate School of Medicine, Yufu, Japan.
Cytokines and high mobility group box chromosomal protein-1 (HMGB-1) play key roles in inflammatory conditions. While hemofiltration has been shown to remove cytokines, removal of cytokines and HMGB-1 by hemofiltration using a polyethersulfone membrane has not been reported. This study aimed to test the hypothesis that the polyethersulfone membrane will achieve higher removal performance for substances including inflammatory cytokines compared to other hemofilters, while retaining low albumin removal capacity. Subjects were eight healthy volunteers. We collected 400 mL each of blood samples into containers with heparin and added 30 mg of lipopolysaccharide to spike cytokines and HMGB-1. After incubation at 39ºC for 12 h, each blood sample was circulated through a hemofiltration circuit with a polyethersulfone hemofilter (2.1 m2 or 1.1 m2 ) at a filtration flow rate of 2 L/h. Measurement samples were collected from arterial, venous, and ultrafiltrate sampling points. Concentrations of cytokines (IL-1ß, IL-4, IL-6, IL-8, IL- 10, and tumor necrosis factor [TNF-a]), HMGB-1, and albumin were determined at each time point (1, 4, 8, 12, and 24h). High sieving coefficients (SCs) above 0.8 were obtained for all cytokines except for TNF-a as well as HMGB-1, whereas the SC for albumin was less than 0.04 with both hemofilters. The hemofilter with a larger membrane area achieved significantly higher clearances for TNF-a and HMGB-1, and slower decreases in SCs over time for IL-1ß, IL-6, IL-8, TNF-a, and albumin. Continuous hemofiltration with a polyethersulfone membrane achieved high efficiency removal of cytokines and HMGB-1, without excessive removal of albumin.
Cytokines and high mobility group box chromosomal protein-1 (HMGB-1) play key roles in inflammatory conditions. While hemofiltration has been shown to remove cytokines, removal of cytokines and HMGB-1 by hemofiltration using a polyethersulfone membrane has not been reported. This study aimed to test the hypothesis that the polyethersulfone membrane will achieve higher removal performance for substances including inflammatory cytokines compared to other hemofilters, while retaining low albumin removal capacity. Subjects were eight healthy volunteers. We collected 400 mL each of blood samples into containers with heparin and added 30 mg of lipopolysaccharide to spike cytokines and HMGB-1. After incubation at 39ºC for 12 h, each blood sample was circulated through a hemofiltration circuit with a polyethersulfone hemofilter (2.1 m2 or 1.1 m2 ) at a filtration flow rate of 2 L/h. Measurement samples were collected from arterial, venous, and ultrafiltrate sampling points. Concentrations of cytokines (IL-1ß, IL-4, IL-6, IL-8, IL- 10, and tumor necrosis factor [TNF-a]), HMGB-1, and albumin were determined at each time point (1, 4, 8, 12, and 24h). High sieving coefficients (SCs) above 0.8 were obtained for all cytokines except for TNF-a as well as HMGB-1, whereas the SC for albumin was less than 0.04 with both hemofilters. The hemofilter with a larger membrane area achieved significantly higher clearances for TNF-a and HMGB-1, and slower decreases in SCs over time for IL-1ß, IL-6, IL-8, TNF-a, and albumin. Continuous hemofiltration with a polyethersulfone membrane achieved high efficiency removal of cytokines and HMGB-1, without excessive removal of albumin.
Related Concept Videos
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Continuing Care
Continuity of a Function
Continuity Equation
The mass flow rate is expressed as:
Continuity Equation