Electron microscopy of silicone irrigation/aspiration tips involved in posterior capsule rupture
John C Hart1, Michael Tanner1, David M Rooney1
1From the Department of Ophthalmology (Hart, Rooney), William Beaumont Hospital, Oakland University/William Beaumont School of Medicine (Hart, Rooney), and the Department of Clinical Engineering (Tanner), William Beaumont Hospital, Royal Oak, Michigan, USA.
Abstract:
We describe 2 cases of posterior capsule rupture caused by Alcon reusable silicone irrigation/aspiration (I/A) tips. Scanning electron microscopy and reflected light microscopy of these tips revealed burring of the inner metallic shafts and tears in the silicone sleeves. A review of surgical video revealed that posterior capsule ruptures caused by the I/A tips occurred when the posterior capsule was aspirated either through the aspiration port or through a tear in the silicone sleeve. Contact of the posterior capsule with the sharp metallic burrs on the inner metal tube can result in posterior capsule rupture.
Insights
Two cases of posterior capsule rupture were linked to Alcon reusable silicone irrigation/aspiration (I/A) tips. Microscopic analysis revealed tip defects, leading to capsule tears during cataract surgery.
Area of Science:
- Ophthalmology
- Surgical Instrumentation
- Cataract Surgery
Background:
- Posterior capsule rupture (PCR) is a potential complication during cataract surgery.
- The integrity of surgical instruments is crucial for patient safety.
Observation:
- Two cases of PCR were associated with reusable Alcon silicone irrigation/aspiration (I/A) tips.
- Microscopic examination showed burring on inner metallic shafts and tears in silicone sleeves of the I/A tips.
Findings:
- PCR occurred when the posterior capsule was aspirated through the I/A tip's aspiration port or a tear in its silicone sleeve.
- Sharp metallic burrs on the inner metal tube of the I/A tips were identified as a cause of posterior capsule rupture.
Implications:
- This highlights the importance of meticulous inspection and maintenance of reusable surgical instruments.
- Potential design improvements or material changes for I/A tips may reduce the risk of PCR.
- Surgeons should be aware of potential instrument-related risks during cataract surgery.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
09:53Micropatterning Transmission Electron Microscopy Grids to Direct Cell Positioning within Whole-Cell Cryo-Electron Tomography Workflows
Published on: September 13, 2021
Related Concept Videos
Overview of Electron Microscopy
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Transmission Electron Microscopy
Immunogold Electron Microscopy
Cryo-electron Microscopy
Design Example: Design of an Irrigation Channel