Assessment of subjective refraction with a clinical adaptive optics visual simulator
Lucía Hervella1, Eloy A Villegas2, Pedro M Prieto2
1Laboratorio de Óptica, Universidad de Murcia, Murcia, Spain; Departamento de Física, Universidad de Murcia, and Voptica SL, Murcia, Spain.
Purpose:
To clinically validate an adaptive optics visual simulator (VAO) that measures subjective refraction and visual acuity.
Setting:
Optics Laboratory, University of Murcia, Murcia, Spain.
Design:
Prospective case series.
Methods:
Using the adaptive optics visual simulator, 2 examiners measured the subjective refraction and visual acuity in healthy eyes of volunteers; 1 examiner also used a trial frame as a gold standard. The interexaminer reproducibility and agreement with the gold standard were estimated using the following statistical parameters: limits of agreement from Bland-Altman analysis, significance between differences (P value), and intraclass correlation coefficient (ICC).
Results:
Seventy-six eyes of 38 volunteers were measured. Interexaminer reproducibility for subjective refraction was excellent (ICC ≥0.96; P > .05), with low 95% confidence interval (CI) values for the power vectors M (spherical equivalent of the given refractive error), J0 (Jackson cross-cylinder, axes at 180 degrees and 90 degrees), and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) (±0.51 diopter [D], ±0.14 D, and ±0.14 D, respectively). No significant differences in subjective refraction and visual acuity were found between the visual simulator and gold standard (P > .05), with 95% CIs for M, J0, and J45 (subjective refraction) of ±0.67 D, ±0.14 D, and ±0.16 D, respectively, and a ±0.10 logarithm of the minimum angle of resolution (visual acuity).
Conclusion:
Subjective refraction results using the adaptive optics visual simulator agreed with those of the gold standard and can be used as the baseline for visual simulation of any optical corneal profile or intraocular lens design for refractive surgery patients.
Related Concept Videos
Assessment of the Gastrointestinal System I: Subjective Data
Health History
The initial step in assessing the GI system is obtaining a comprehensive health history. This includes inquiring about the patient's history or presence of problems...
Assessment of the Cardiovascular System I: Subjective Data
Initial Enquiry
Ask the patient about their primary concern and thoroughly explore all reported symptoms.
Medical History
Investigate past illnesses affecting the cardiovascular system, such as angina, anemia, rheumatic fever, congenital heart disease, stroke, thrombophlebitis, dysrhythmias, varicosities
Inquire about symptoms...
Angina III: Clinical Manifestations and Assessment
Cable Subjected to Its Own Weight
A generalized loading function is employed to analyze a cable subjected to its own weight. This function considers the force acting along the cable's arc length rather than its projected length, providing a more accurate...
Influence of Earth's Curvature and Atmospheric Refraction on Leveling
Cable Subjected to Concentrated Loads