Usefulness of slice encoding for metal artifact correction (SEMAC) technique for reducing metal artifacts after total
Ahmed Jawhar1, Miriam Reichert2, Michael Kostrzewa3
1Department of Orthopaedics and Trauma Surgery, University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany. Ahmed.Jawhar@umm.de.
Purpose:
To evaluate the usefulness of a novel MRI sequence strategy in the assessment of the periprosthetic anatomical structures after primary total knee arthroplasty.
Methods:
Two MR sequences were retrospectively compared for the imaging of 15 patients with implanted cruciate-retaining/fixed-bearing TKAs (DePuy, PFC Sigma): a slice encoding sequence for metal artifact correction (SEMAC) and a standard sequence. Images were acquired on a 1.5-T system. The degree of artifact reduction was assessed using several qualitative (Likert-type scale) (artifact size, distorsion, blur, image quality, periprosthetic bone, posterior cruciate ligament, lateral collateral ligament, medial collateral ligament, patella tendon, popliteal vessels) and quantitative (artifact volume, Insall-Salvati index, length of patella/tendon, prosthesis dimensions) parameters by blinded reads performed by four investigators. The SEMAC sequences were statistically compared with the standard sequence using Wilcoxon test. Additionally, the intraclass correlation coefficient (ICC) for interobserver agreement was calculated.
Results:
Higher levels of blurring were found with SEMAC compared to standard sequences (p < 0.001). All other qualitative parameters improved significantly with the application of SEMAC. In comparison with conventional sequences, the artifact volume was reduced by 59% utilizing SEMAC. Thus, the artifact reduction improved the precision of measurements such as Insall-Salvati index and length of patella/tendon (p < 0.001). The dimension of the tibial component (Ti alloy/polyethylene) revealed accurate values with both MRI sequences. A sufficient interobserver agreement among all readers was found with SEMAC, qualitatively ICC 0.9 (range 0.8-1) as well as quantitatively ICC 0.95 (range 0.92-0.98).
Conclusions:
SEMAC effectively reduces artifacts caused by metallic implants after total knee arthroplasty relative to standard imaging. This allows for an improved assessment of periprosthetic anatomical structures. This might enable an improved detectability of postoperative complications in the future.
Level Of Evidence:
Diagnostic Study Level III.
Related Concept Videos
Alkali Metals
Table 1: Properties of the alkali metals
Bonding in Metals
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Properties of Transition Metals
Theory of Metallic Conduction
In this theory, Newton's second law of motion is used to determine the acceleration of an electron in the presence of an applied electric field. Then, its velocity is expressed via this acceleration.
An electron moves through the crystal, containing positive ions,...


