Separating High-Z Oral Contrast From Intravascular Iodine Contrast in an Animal Model Using Dual-Layer Spectral CT
Todd C Soesbe1, Matthew A Lewis1, Khaled Nasr1
1Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines BLVD, Dallas, TX 75390-8896.
Rationale And Objectives:
To show that water and iodine two-material decomposition images from dual-layer dual-energy spectral X-ray computed tomography (DECT) can be used to separate intravascular iodine contrast from simultaneously administered oral tantalum, tungsten, or rhenium contrast in an animal model.
Materials And Methods:
In this Institutional Animal Care and Use Committee approved study, four female Fischer rats were given simultaneous intravenous and oral X-ray computed tomography contrast. Intravenous iodine contrast was administered via tail vein injection. Oral barium, tantalum, tungsten, or rhenium contrast was administered via gavage. The animals were imaged on a dual-layer DECT system at 120 kVp. Water and iodine two-material decomposition images (water equivalent and iodine equivalent images) were used for qualitative analysis. Computer simulations were performed using a customized DECT simulator to better understand why certain high-Z elements disappear in the iodine equivalent images and what is the theoretical range of elements with this property.
Results:
The iodine and barium contrast appeared only in the iodine equivalent images and could not be differentiated from each other. However, the tantalum, tungsten, and rhenium contrast only appeared in the water equivalent images. This allowed iodine contrast in the bowel wall to be easily segmented from tantalum, tungsten, and rhenium contrast in the bowel lumen. Simulations confirmed that certain high-Z elements will have pixel values of ≤0 mg iodine/mL in the iodine equivalent images due to a K-edge effect associated with DECT systems.
Conclusions:
Dual-layer DECT can separate iodine from certain high-Z elements using water equivalent and iodine equivalent images with an increased element range compared to other DECT systems. This K-edge effect could promote the development and approval of new high-Z contrast agents for DECT.
More Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Tonicity in Animals
Theories of Dissolution: Diffusion Layer Model
This process starts with a thin layer, saturated with the drug, forming at the interface between the solid and liquid. The solute then diffuses from this layer into the main solution. The Noyes-Whitney equation suggests that the rate of dissolution relies on the diffusion...
Oral Cavity
Teeth: The teeth are the hardest structures in our bodies. Humans have two sets of teeth throughout their lifetime: deciduous (baby) teeth and permanent teeth. Each tooth consists of several parts: the crown (visible part), the root (embedded in the jaw), enamel (hard outer...
Electrophilic Aromatic Substitution: Fluorination and Iodination of Benzene
Layers of the Epidermis
Stratum Basale
Stratum basale, also known as the stratum germinativum, is the deepest layer of the epidermis. It is composed of a single layer of actively dividing cells called basal cells or basal keratinocytes. These cells constantly undergo cell division to replenish the upper layers of the epidermis. Additionally, melanocytes, which...