Late Onset Valve Dislocation of the Edwards Intuity Rapid-Deployment Bioprosthesis
Max Wacker1, Ingo Slottosch1, Maximilian Scherner1
1Department of Cardiothoracic Surgery, University Hospital of Magdeburg, Magdeburg, Germany.
Abstract:
Over the last decade, aortic valve replacement with rapid-deployment biologic bioprostheses has become a common alternative to the use of mechanical or biologic stented valves for high-risk patients. A 63-year-old patient underwent uncomplicated valve replacement with the Edwards Intuity valve (Edwards Lifesciences, Irvine, CA). Two months postoperatively, the patient had progressive dyspnea. Echocardiography showed a dislocated aortic valve reaching into the left ventricular outflow tract and impairing the anterior mitral valve leaflet. Both valves were replaced by a conventional stented bioprosthesis. This case report describes the first case of valve migration for the Edwards Intuity Elite rapid-deployment aortic valve and discusses possible explanations.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...