Adjustable-length loop cortical button versus interference screw fixation in quadriceps tendon anterior cruciate
Roman Frederik Karkosch1, Max Ettinger1, Samuel Bachmaier2
1Department of Orthopedic Surgery, Hannover Medical School, Anna-von-Borries Straße 1-7, 30625 Hanover, Germany.
Background:
This biomechanical cadaveric in vitro study aimed to evaluate and compare the dynamic elongation behavior and ultimate failure strength of tibial adjustable-length loop cortical button versus interference screw fixation in quadriceps tendon-based anterior cruciate ligament reconstruction.
Methods:
Sixteen human quadriceps tendons were harvested and fixed into porcine tibiae using either biodegradable interference screw (n = 8) or adjustable loop device (n = 8) fixation. An acrylic block was utilized for femoral adjustable loop device fixation for both groups. All constructs were precycled for 10 times at 0.5 Hz and manually retensioned before tested in position and force control mode each for 1000 cycles at 0.75 Hz according to in vitro loading conditions replicating the in vivo ACL environment. Subsequently, an ultimate failure test at 50 mm/min was performed with mode of failure noted.
Findings:
Tibial IS fixation showed no statistically significant differences in the initial (-0.46 vs. -0.47 mm; P = 0.9780), dynamic (2.18 mm vs. 2.89 mm; P = 0,0661), and total elongation (1.72 mm vs. 2.42 mm; P = 0,0997) compared to adjustable loop device fixation. The tibial button fixation revealed an increased ultimate failure load (743.3 N vs. 606.3 N; P = 0.0027), while stiffness was decreased in comparison to screw fixation (133.2 N/mm vs. 153.5 N/mm; P = 0,0045).
Interpretation:
Anterior cruciate ligament reconstruction for quadriceps tendon graft using a tibial adjustable-length loop cortical button provides for comparable dynamic stabilization of the knee with increased ultimate failure load at decreased stiffness compared to screw fixation.
More Related Videos
10:32Athymic Rat Model for Evaluation of Engineered Anterior Cruciate Ligament Grafts
Published on: March 26, 2015
06:28Anterior Cruciate Ligament Transection and Synovial Fluid Lavage in a Rodent Model to Study Joint Inflammation and Posttraumatic Osteoarthritis
Published on: September 2, 2025
Related Concept Videos
Interference and Diffraction
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
The Anchoring-and-Adjustment Heuristic
Adjusting a Traverse