Related Experiment Video
Updated: Feb 3, 2026

Combining 3D-Printing and Electrospinning to Manufacture Biomimetic Heart Valve Leaflets
Published on: March 23, 2022
Pre-procedural fit-testing of TAVR valves using parametric modeling and 3D printing
Ahmed Hosny1, Joshua D Dilley2, Tatiana Kelil3
1Dana-Farber Cancer Institute, Harvard Medical School, USA.
Background:
Successful transcatheter aortic valve replacement (TAVR) requires an understanding of how a prosthetic valve will interact with a patient's anatomy in advance of surgical deployment. To improve this understanding, we developed a benchtop workflow that allows for testing of physical interactions between prosthetic valves and patient-specific aortic root anatomy, including calcified leaflets, prior to actual prosthetic valve placement.
Methods:
This was a retrospective study of 30 patients who underwent TAVR at a single high volume center. By design, the dataset contained 15 patients with a successful annular seal (defined by an absence of paravalvular leaks) and 15 patients with a sub-optimal seal (presence of paravalvular leaks) on post-procedure transthoracic echocardiogram (TTE). Patients received either a balloon-expandable (Edwards Sapien or Sapien XT, n = 15), or a self-expanding (Medtronic CoreValve or Core Evolut, n = 14, St. Jude Portico, n = 1) valve. Pre-procedural computed tomography (CT) angiograms, parametric geometry modeling, and multi-material 3D printing were utilized to create flexible aortic root physical models, including displaceable calcified valve leaflets. A 3D printed adjustable sizing device was then positioned in the aortic root models and sequentially opened to larger valve sizes, progressively flattening the calcified leaflets against the aortic wall. Optimal valve size and fit were determined by visual inspection and quantitative pressure mapping of interactions between the sizer and models.
Results:
Benchtop-predicted "best fit" valve size showed a statistically significant correlation with gold standard CT measurements of the average annulus diameter (n = 30, p < 0.0001 Wilcoxon matched-pairs signed rank test). Adequateness of seal (presence or absence of paravalvular leak) was correctly predicted in 11/15 (73.3%) patients who received a balloon-expandable valve, and in 9/15 (60%) patients who received a self-expanding valve. Pressure testing provided a physical map of areas with an inadequate seal; these corresponded to areas of paravalvular leak documented by post-procedural transthoracic echocardiography.
Conclusion:
We present and demonstrate the potential of a workflow for determining optimal prosthetic valve size that accounts for aortic annular dimensions as well as the active displacement of calcified valve leaflets during prosthetic valve deployment. The workflow's open source framework offers a platform for providing predictive insights into the design and testing of future prosthetic valves.
Related Concept Videos
Induced-fit Model
Enzymes exhibit substrate specificity, meaning that they can only bind to certain substrates. This is mainly determined by the shape and chemical...
Statistical Methods to Analyze Parametric Data: Student t-Test and Goodness-of-Fit Test
The Student's t-test is a statistical test that examines if there is a statistically significant difference between the means of two groups. This test is instrumental when dealing with...
Expected Frequencies in Goodness-of-Fit Tests
Handwashing II: Pre-procedure and Initial Procedure Steps
The hand washing procedure itself includes the following steps. First, cover cuts, if any, on hands with a waterproof dressing. Cuts and abrasions can become contaminated with bacteria hindering the ability to clean the area thoroughly. In addition, repeated hand washing can worsen an injury. The nails must be short and clean, without nail...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Cardiac Catheterization I: Pre-Procedure Overview