The effectiveness of additional lead-shielding drape and low pulse rate fluoroscopy in protecting staff from scatter
Yoshiaki Morishima1,2, Koichi Chida3, Yoshiaki Katahira4
1Department of Radiology, Tohoku Medical and Pharmaceutical University Hospital, 1-12-1 Fukumuro, Miyagino-ku, Sendai, 983-8512, Japan. morishima@med.tohoku.ac.jp.
Purpose:
Cardiac resynchronization therapy (CRT) often requires a long fluoroscopic time and protection from scatter radiation. This study reports on scatter radiation levels during CRT, with and without additional shielding, and using standard or low pulse rate fluoroscopy.
Materials And Methods:
Additional lead-shielding drape (0.35-mm lead equivalent) was used on the left side of the table and pulsed fluoroscopy was performed at rates of 10 pulses/s (usual rate) and 7.5 pulses/s (low pulse rate). Fluoroscopy scatter radiation was measured for both pulse rates using an acrylic phantom with a radiation survey meter, both with and without the additional lead-shielding drape.
Results:
With the additional lead-shielding drape, the fluoroscopy scatter radiation was reduced by 74.3% at 10 pulses/s and 78.6% at 7.5 pulses/s. If the fluoroscopy was changed from 10 pulses/s to 7.5 pulses/s, the scattered radiation at the primary physician's position was reduced by 24.0%. The combined use of additional shielding drape and low pulse rate fluoroscopy reduced scatter radiation by over 80%.
Conclusion:
Additional lead-shielding drape and low pulse rate fluoroscopy are effective in reducing the scattered radiation dose to physicians and nurses during CRT.
Related Concept Videos
Biological Effects of Radiation
Increased pulse rate
Many factors can elevate the risk of developing tachycardia. These include advanced age, a family history of arrhythmias, and an...
Decreased pulse rate
There are specific risk factors that can elevate the likelihood of developing bradycardia. Advanced age is a significant factor, with...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Conjugate Addition (1,4-Addition) vs Direct Addition (1,2-Addition)
Conjugate addition results in a thermodynamically stable product. The reaction retains the stronger C=O bond at the expense of the weaker C=C π bond. The process is slow as the β carbon is less electrophilic than the carbonyl carbon.
Direct addition products are...
Pulse
The pulse serves as a clinical...