Precision and reliability of liver iodine quantification from spectral detector CT: evidence from phantom and patient
Nils Große Hokamp1,2,3, Nuran Abdullayev4, Thorsten Persigehl4
1Institute for Diagnostic and Interventional Radiology, University Hospital Cologne, Cologne, Germany. nils.grosse-hokamp@uk-koeln.de.
Objective:
To comprehensively assess precision, reproducibility, and repeatability of iodine maps from spectral detector CT (SDCT) in a phantom and in patients with repetitive examination of the abdomen.
Methods:
Seventy-seven patients who underwent examination two (n = 52) or three (n = 25) times according to clinical indications were included in this IRB-approved, retrospective study. The anthropomorphic liver phantom and all patients were scanned with a standardized protocol (SSDE in patients 15.8 mGy). In patients, i.v. contrast was administered and portal venous images were acquired using bolus-tracking technique. The phantom was scanned three times at three time points; in one acquisition, image reconstruction was repeated three times. Region of interest (ROI) were placed automatically (phantom) or manually (patients) in the liver parenchyma (mimic) and the portal vein; attenuation in conventional images (CI [HU]) and iodine map concentrations (IM [mg/ml]) were recorded. The coefficient of variation (CV [%]) was used to compare between repetitive acquisitions. If present, additional ROI were placed in cysts (n = 29) and hemangioma (n = 29).
Results:
Differences throughout all phantom examinations were < 2%. In patients, differences between two examinations were higher (CV for CI/IM: portal vein, 2.5%/3.2%; liver parenchyma, -0.5%/-3.0% for CI/IM). In 80% of patients, these differences were within a ± 20% limit. Differences in benign liver lesions were even higher (68% and 38%, for CI and IM, respectively).
Conclusions:
Iodine maps from SDCT allow for reliable quantification of iodine content in phantoms; while in patients, rather large differences between repetitive examinations are likely due to differences in biological distribution. This underlines the need for careful clinical interpretation and further protocol optimization.
Key Points:
• Spectral detector computed tomography allows for reliable quantification of iodine in phantoms. • In patients, the offset between repetitive examinations varies by 20%, likely due to differences in biological distribution. • Clinically, iodine maps should be interpreted with caution and should take the intra-individual variability of iodine distribution over time into account.
Related Concept Videos
The Evidence for Evolution
Reliability and Validity
Data Reporting and Recording
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II