Accuracy of using high-energy prompt gamma to verify proton beam range with a Compton camera: A Monte Carlo
Hsuan-Ming Huang1, Meei-Ling Jan2
1Institute of Medical Device and Imaging, College of Medicine, National Taiwan University, No.1, Sec 1, Jen Ai Rd., Zhongzheng Dist., Taipei City 100, Taiwan.
Abstract:
Prompt gamma (PG) rays emitted during proton therapy has been used for proton range verification. Because high-energy PG emission is well correlated to the Bragg peak (BP), high-energy PG rays are well-suited for proton range verification. However, the low production and detection of high-energy PG rays often lead to inaccurate BP position estimates. The aim of this study is to improve the BP position estimates obtained from high-energy PG rays. We propose a BP position estimation method based on the local maximum closest to the distal fall-off region. We present the results of Monte Carlo simulations in which a water phantom was irradiated with a proton beam. Our results show that the BP position estimated from the 6.13 MeV PG rays can be improved using the proposed position estimation method. Moreover, the 6.92 and 7.12 MeV PG rays can be used for predicting the BP position. However, the accuracy of the BP position estimation decreases with decreasing tissue oxygen levels. We also found that the subtraction of the PG images of 6.13 MeV from those of 6.92 and 7.12 MeV can be used to predict the BP position with a mean accuracy of < 2 mm. The accurate estimation of the BP position can be achieved using different high-energy PG rays, but factors including position estimation, irradiated tissue and event selection should be carefully taken into account.
Related Concept Videos
Nuclear Binding Energy
Improving Translational Accuracy
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Uncertainty in Measurement: Accuracy and Precision
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Light as Energy
Photons
A photon is a discrete electromagnetic particle or bundle of energy. Photons are characterized by their frequency, wavelength, and amplitude, similar to the properties of a wave. Waves with higher frequencies transmit more energy and have shorter wavelengths than longer wavelengths that transmit...