Precision of Dual-Energy X-Ray Absorptiometry Reflection Scans in Muscular Athletes
Grant M Tinsley1, Michael Lane Moore1, Austin J Graybeal1
1Department of Kinesiology & Sport Management, Texas Tech University, Lubbock, TX, USA.
Background:
Offset scanning procedures, such as reflection scanning, allow for dual-energy x-ray absorptiometry (DXA) body composition assessment of individuals who are too broad for standard scanning dimensions. However, limited information is available concerning the precision of this procedure, particularly in athletes.
Methodology:
Twenty-seven muscular athletes (n = 17 males, BMI: 28.8 ± 2.0 kg/m2, DXA body fat: 12.5 ± 2.7%; n = 10 females, BMI: 22.8 ± 1.6 kg/m2, DXA body fat: 19.2 ± 3.4%) underwent consecutive DXA scans on a GE Lunar Prodigy scanner using the reflection scanning technique. The fully automated output was obtained for each scan, and an additional version of each scan's output was saved after manual adjustment of regions of interest (ROI). Metrics of reliability and precision were calculated for total and regional body mass (BM), lean mass (LM), fat mass (FM) and bone mineral content (BMC). These metrics included the precision error (PE), least significant change, ΔMean, technical error of measurement, intraclass correlation coefficient, smallest worthwhile effect and minimum difference considered real.
Results:
Reflection scanning produced small errors for BM (PE: ∼0.5%), LM (PE < 1%) and BMC (PE: ∼1.2%), with larger errors observed for total FM (PE: ∼3%). Manual ROI adjustment produced lower errors for total BM, LM, FM, and BMC, as well as lower errors for most regional estimates. The utilization of automated ROI revealed concerns unique to reflection scanning, including unnecessary estimation of trunk composition, which appreciably increased error in this region. Regional estimates produced higher errors for all variables as compared to whole-body estimates, although which regions produced the highest errors differed between BM, LM, FM, and BMC.
Conclusions:
Reflection scanning allows DXA body composition assessment in individuals exceeding traditional scanning dimensions, including broad athletes. Although this procedure introduces error, it may be minimized through manual adjustment of ROIs and consistency of analysis methods.
Related Concept Videos
The Muscular System
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Light as Energy
Photons
A photon is a discrete electromagnetic particle or bundle of energy. Photons are characterized by their frequency, wavelength, and amplitude, similar to the properties of a wave. Waves with higher frequencies transmit more energy and have shorter wavelengths than longer wavelengths that transmit...
Leaky Scanning
What is Energy?
Free Energy