Investigation of the quantitative accuracy of low-dose amyloid and tau PET imaging
Ying-Hwey Nai1,2, Shoichi Watanuki3, Manabu Tashiro3
1Division of Radiation Informatics for Medical Imaging, Graduate School of Biomedical Engineering, Tohoku University, Sendai, Japan.
Abstract:
With the increasing incidence of dementia worldwide, the frequent use of amyloid and tau positron emission tomography imaging requires low-dose protocols for the differential diagnoses of various neurodegenerative diseases and the monitoring of disease progression. In this study, we investigated the feasibility to reduce the PET dose without a significant loss of quantitative accuracy in 3D dynamic row action maximum likelihood algorithm-reconstructed PET images using [11C]PIB and [18F]THK5351. Eighteen cognitively normal young controls, cognitively normal elderly controls, and patients with probable Alzheimer's disease (n = 6 each), were included. Reduced doses were simulated by randomly sampling half and quarter of the full counts in list mode data for one independent realization at each simulated dose. Bias was evaluated between the reduced dose from the full dose of standardized uptake value ratio (SUVR), distribution volume ratio (DVR) from reference Logan, and non-displaceable binding potential (BPND) from simplified reference tissue model (SRTM). DVR yielded the least bias at low dose compared to SUVR and BPND, and thus, is highly recommended. The dose of [18F]THK5351 and [11C]PIB can be reduced to a quarter of the full dose using DVR for evaluation, whereas the dose can only be reduced to half and a quarter of the full dose for [18F]THK5351 and [11C]PIB using SUVR. BPND showed inconsistent trend and large bias at low dose. The feasibility of dose reduction was dependent on the selected parameters of interest, reconstruction algorithms, reference regions, and to a lesser degree by motion effects.
Related Concept Videos
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision

![Visualization and Quantification of Brown and Beige Adipose Tissues in Mice using [18F]FDG Micro-PET/MR Imaging](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F62460.jpg&w=3840&q=50)