Hemolysis risk after packed red blood cells transfusion with infusion pumps
Ana Maria Miranda Martins Wilson1, Maria Angélica Sorgini Peterlini2, Mavilde da Luz Gonçalves Pedreira2
1Universidade de São Paulo, Escola de Enfermagem, São Paulo, SP, Brazil.
Objective:
To evaluate the hemolysis biomarkers of packed red blood cells transfused by two different linear peristaltic infusion pumps at two infusion rates.
Method:
An experimental and randomized study was designed simulating the clinical practice of transfusion. Two linear peristaltic infusion pumps from different manufactures were studied in triplicate at 100 mL/h and 300mL/h infusion rates. The chosen hemolysis biomarkers were total hemoglobin, free hemoglobin, hematocrit, potassium and degree of hemolysis. They were analyzed before and after each infusion.
Results:
Potassium showed statistically significant variations in all scenarios of the experiment (P<0.010). In a separated analysis, potassium increased mainly at 300mL/h rate (P=0.021) and free hemoglobin had significant variation when comparing infusion pumps from different manufacturers (P=0.026). Although hematocrit, total hemoglobin and degree of hemolysis had increased after infusion, no statistically significance variations were identified.
Conclusions:
Hemolysis risk induced by a linear peristaltic infusion pump was identified by an increase in free hemoglobin and potassium markers. As the potassium biomarker is often increased in aged packed red blood cells, we do not recommend using them in this scenario. Additional studies should be performed about other markers and using larger samples in order to reinforce the transfusion practice in nursing.
More Related Videos
11:12Ex Vivo Red Blood Cell Hemolysis Assay for the Evaluation of pH-responsive Endosomolytic Agents for Cytosolic Delivery of Biomacromolecular Drugs
Published on: March 9, 2013
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
Related Concept Videos
Blood Transfusion
Blood Transfusion Overview
A blood transfusion is a medical procedure used to replace blood lost due to injury, surgery, or to treat conditions such as anemia or cancer. During a transfusion, donor blood is...
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...
Relative Risk
Pumped Concrete
For direct-acting pumps, the concrete enters the pump via the inlet valve under the action of gravity and suction created by the movement of the piston. This concrete is then forced into the pipeline and out through the outlet valve by the forward movement...