Comparison of 577-nm Multispot and Standard Single-Spot Photocoagulation for Diabetic Retinopathy
Renato M Passos1, José Belucio-Neto1, Camilla O Xavier1
1Department of Ophthalmology, Federal University of São Paulo (UNIFESP), São Paulo, Brazil.
Objective:
To compare two different laser strategies of panretinal photocoagulation for diabetic retinopathy.
Methods:
Single-center, randomized study including 41 eyes treated with 577-nm multispot laser with a 20-ms pulse duration (group 1) or a 532-nm single-spot laser with a 100-ms pulse duration (group 2). The outcomes included best-corrected visual acuity (BCVA) and imaging changes at baseline, 6 and 12 months, laser parameters, and results of subjective pain analysis.
Results:
At 12 months, the treatments did not differ significantly in BCVA, central retinal thicknesses (CRTs), improved macular edema, vitreomacular interface changes, patient-reported pain scores, or angiographic responses. Group 1 had significantly fewer treatment sessions but used more laser spots (p < 0.001).
Conclusion:
The multispot laser required fewer applications with more spots delivered to compensate for lower fluency, showing similar patient tolerance to single-spot laser. Both groups maintained the initial visual acuities and CRTs; about 50% of cases had vitreomacular interface changes and improved macular edema, with similar angiographic improvements after 12 months.
Related Concept Videos
Measurement: Standard Units
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Pathophysiology of Diabetes
Type 1 diabetes is characterized by autoimmune-mediated destruction of pancreatic β cells, with environmental factors potentially triggering this process in genetically susceptible individuals. Despite many not having a family history, certain genes increase susceptibility,...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Standard Enthalpy of Formation
Standard Electrode Potentials