Validation of a virtual intracorporeal suturing simulator
Related Experiment Video
Updated: Feb 3, 2026

Developing a Virtual Reality Video Game to Simulate Rip Currents
Published on: July 16, 2020
Yaoyu Fu1, Lora Cavuoto2, Di Qi3
1Department of Industrial and Systems Engineering, University at Buffalo, Buffalo, NY, 14260, USA.
Background:
Intracorporeal suturing is one of the most important and difficult procedures in laparoscopic surgery. Practicing on a FLS trainer box is effective but requires large number of consumables, and the scoring is somewhat subjective and not immediate. A virtualbasic laparoscopic skill trainer (VBLaST©) was developed to simulate the five tasks of the FLS Trainer Box. The purpose of this study is to evaluate the face and content validity of the VBLaST suturing simulator (VBLaST-SS©).
Methods:
Twenty-five medical students and residents completed an evaluation of the simulator. The participants were asked to perform the standard intracorporeal suturing task on both VBLaST-SS© and the traditional FLS box trainer. The performance scores on each system were calculated based on time (s), deviations to the black dots (mm), and incision gap (mm). The participants were then asked to finish a 13-item questionnaire with ratings from 1 (not realistic/useful) to 5 (very realistic/useful) regarding the face validity of the simulator. A Wilcoxon signed rank test was performed to identify differences in performance on the VBLaST-SS© compared to that of the traditional FLS box trainer.
Results:
Three questions from the face validity questionnaire were excluded due to lack of response. Ratings to 8 of the remaining 10 questions (80%) averaged above 3.0 out of 5. Average intracorporeal suturing completion time on the VBLaST-SS© was 421 (SD = 168 s) seconds compared to 406 (175 s) seconds on the box trainer (p = 0.620). There was a significant difference between systems for the incision gap (p = 0.048). Deviation in needle insertion from the black dot was smaller for the box trainer than the virtual simulator (1.68 vs. 7.12, p < 0.001).
Conclusion:
Participants showed comparable performance on the VBLaST-SS© and traditional box trainer. Overall, the VBLaST-SS© system showed face validity and has the potential to support training for the suturing skills.
Related Concept Videos
Reliability and Validity
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Principle of Virtual Work: Problem Solving
To apply the principle of virtual work,...