In vitro biomechanical evaluation of a monocoque plate-spacer construct for cervical open-door laminoplasty
Yukoh Ohara1,2, Takeshi Hara2, Alejandro A Espinoza Orías3
1Center for minimally invasive spinal surgery, Shin-yurigaoka General Hospital, Kawasaki, Japan.
Purpose:
To evaluate biomechanical differences between two surgical procedures for cervical open-door laminoplasty using human cadaveric spines.
Methods:
Twenty-four cervical vertebrae (C4-C6) from eight fresh-frozen human cervical spines were subjected to mechanical testing after being instrumented for open-door laminoplasty using a newly designed plate-spacer device with a monocoque structure (plate-spacer group; n = 12) or by conventional miniplate-alone fixation (miniplate group; n = 12). Cantilever bending testing was performed by applying a compressive load in the cranio-caudal direction to the base of the spinous process of the reconstructed laminar arch constructs until failure and strength and stiffness of the laminar arch were determined. The results are presented as mean ± standard deviation.
Results:
The plate-spacer group was approximately twice as strong as the miniplate group (27.6 ± 16.5 N vs. 13.5 ± 7.3 N, p < 0.05). Stiffness in the plate-spacer group exhibited the same trend (19.6 ± 9.3 N/mm vs. miniplate group: 11.4 ± 6.9 N/mm, p < 0.05).
Conclusion:
The fixation with the monocoque plate-spacer construct for open-door laminoplasty provided higher structural properties when compared against the plate-alone fixation. The spacer in the plate-spacer construct appears to contribute by preventing large deformations of the laminar arch caused by bending in cranio-caudal direction. Future studies will be required to investigate stress/strain distribution in the laminar arch constructs.
Related Concept Videos
Construction of Root Locus
For positive gain values, the root locus exists on the real axis to the left of an odd number of finite open-loop poles or zeros. The root locus starts at the open-loop poles and traces the paths of the closed-loop poles as the gain...
Construction of Frequency Distribution
First, make a table with two columns—one with the title of the data that needs to be organized, and the other column for frequency. [Draw a third column for tally marks if needed]. Then, take a look at the items given in the data set and decide if an ungrouped frequency distribution table or a grouped frequency distribution table would be more suitable. If there are large sets of different values, then it is...
Bode Plots Construction
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Electric Field of Parallel Conducting Plates
Consider a cross-section of a thin, infinite conducting plate having a positive charge. For such a large thin plate, as the thickness of the plate tends to zero, the positive charges lie on the plate's two large faces. Without an external electric field, the...
Fluid Pressure over Curved Plate of Constant Width