Related Experiment Video
Updated: Feb 3, 2026

Isolation of Mouse Interstitial Valve Cells to Study the Calcification of the Aortic Valve In Vitro
Published on: May 10, 2021
Performance changes of venous valves following tissue treatment with novel in vitro system
Garrett Easson1, Megan Laughlin1, Hanna Jensen1
11 Department of Biomedical Engineering, University of Arkansas, Fayetteville, AR, USA.
Objectives:
The purpose of this study is to test venous valve performance and identify differences between native tissue and replacement devices developed with traditional tissue treatment methods using a new in vitro model with synchronized hemodynamic parameters and high-speed valve image acquisition.
Methods:
An in vitro model mimicking the venous circulation to test valve performance was developed using hydrostatic pressure driven flow. Fresh and glutaraldehyde-treated vein segments were placed in the setup and opening/closing of the valves was captured by a high-speed camera. Hemodynamic data were obtained using synchronized hardware and virtual instrumentation.
Results:
Geometric orifice area and opening/closing time of the valves was evaluated at the same hemodynamic conditions. A reduction in geometric orifice area of 27.2 ± 14.8% (p < 0.05) was observed following glutaraldehyde fixation. No significant difference in opening/closing time following chemical fixation was observed.
Conclusions:
The developed in vitro model was shown to be an effective method for measuring the performance of venous valves. The observed decrease in geometric orifice area following glutaraldehyde treatment indicates a decrease in flow through the valve, demonstrating the consequences of traditional tissue treatment methods.
Related Concept Videos
Drug Product Performance: In Vitro–In Vivo Correlation
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Venous Thrombosis I: Introduction
Mitral Valve Prolapse I: Introduction
Venous Thrombosis III: Interprofessional Care