Evaluation of valve function in antireflux biliary metal stents
Chang-Il Kwon1, Jong Pil Moon2, Ho Yun2
1Digestive Disease Center, CHA Bundang Medical Center, CHA University, Seongnam, Republic of Korea.
Background:
To overcome duodenobiliary reflux induced by biliary stents, antireflux valve (ARV) biliary stents have been developed and showed improvement in stent patency. However, negative study results have also been reported because stent patency may be decreased by the malfunction of ARV itself. Given such mixed results, the true efficacy of ARV remains unknown and the mechanism of its dysfunction needs to be clearly elucidated. The aim of this study was to investigate the exact mechanism of ARV dysfunction using in vitro phantom models.
Methods:
Two experimental models were designed to evaluate two important environmental factors suspected to cause ARV malfunction, i.e. bile flow and pH. Three types of ARV metal stents from different companies were used for the experiments: a funnel type ARV, a windsock type ARV, and a wine glass-shaped ARV. Ten stents of each type were tested (five stents in the bile flow phantom model, and another five stents in the duodenal pH environmental model). To determine ARV malfunction, ARV-induced flow resistance was measured using a custom-made testing device. All stents from the two models were removed every 2 weeks for 12 weeks after stent insertion and were evaluated on morphological and functional changes of the ARV.
Results:
Only ARV of wine glass-shaped ARV was morphologically changed due to silicone bond detachment in the bile flow model. All types of ARV were morphologically changed in the pH model. The morphological changes of ARV influenced the flow resistance. The antegrade pressure gradients were increased over time in the pH model (p < 0.05).
Conclusions:
Morphological change of the ARVs may induce dysfunction of ARV metal stents, which is mainly due to duodenal pH environment. In the future, development of new ARV that is not affected by duodenal environmental factors can be expected to improve stent patency.
More Related Videos
06:55Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Bonding in Metals
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Alkali Metals
Table 1: Properties of the alkali metals
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Properties of Transition Metals