Monte Carlo and experimental high dose rate 192Ir brachytherapy dosimetry with microDiamond detectors
Giulio Rossi1, Mark Gainey1, Benedikt Thomann1
1Division of Medical Physics, Department of Radiation Oncology, Medical Center, Faculty of Medicine, University of Freiburg, German Cancer Consortium (DKTK) Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg, Germany.
Abstract:
The purpose of this study was to investigate the suitability of the microDiamond detector (mDD) type 60019 (PTW-Freiburg, Germany) for radial dose function measurements with High Dose Rate (HDR) 192Ir brachytherapy sources. An HDR 192Ir source model mHDR v2r (Nucletron BV, an Elekta company, The Netherlands) was placed at the centre of a MP3 water phantom (PTW-Freiburg, Germany) within a 4F needle. Three mDDs were employed to measure the radial dose function of the source by acquiring profiles along the source transverse axis. Meanwhile, the experimental setup was simulated using the Monte Carlo (MC) code MCNP6.1™ (Los Alamos National Laboratory, USA) to calculate phantom-size, absorbed-dose energy dependence and volume averaging correction factors. After applying the correction factors, the radial dose function gL(r) for the line source approximation was calculated as defined in the TG-43 formalism at radial distances from 0.5cm to 10cm and compared to the consensus gL(r) (ESTRO and AAPM). The percentage differences to the consensus gL(r) for all the three mDDs were from -2.3% to +1.4% for distances r≤5cm and -6.2% to +2.6% for larger distances. These results indicate the suitability of the mDD for HDR brachytherapy measurements when all required corrections are applied.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration