Accuracy and reliability of knee goniometry methods
Graeme Ethan Hancock1, Tracey Hepworth2, Kevin Wembridge2
1Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield, UK. graemehancock@me.com.
Journal of Experimental Orthopaedics
|October 21, 2018
Summary
For accurate knee range of motion measurement, a digital inclinometer is best. Traditional visual estimation and short arm goniometers are least accurate for assessing knee angles.
Area of Science:
- Orthopedics
- Biomechanics
- Medical Measurement Technology
Background:
- Accurate measurement of knee range of motion is crucial for clinical examination and post-operative assessment.
- Various tools exist for measuring knee angles, including goniometers, smartphone applications, and digital inclinometers.
- Establishing the minimum significant difference between measurement methods is vital for monitoring changes effectively.
Purpose of the Study:
- To evaluate the reliability and minimum significant difference of different methods for measuring knee range of motion.
- To compare visual estimation, short and long arm goniometers, a smartphone application, and a digital inclinometer.
- To determine the most accurate method for knee angle assessment.
Main Methods:
- Knee angles were measured by three distinct users (orthopaedic surgeon, surgical trainee, physiotherapist).
- Five methods were employed across six knees, assessing three angles plus full active flexion and extension.
- Measurements were repeated with users blinded to previous results, totaling 300 measurements.
Main Results:
- All methods demonstrated high inter-rater and intra-rater reliability (all > 0.99 and > 0.98, respectively).
- The digital inclinometer exhibited the highest accuracy with a minimum significant difference of 6°.
- Visual estimation and short arm goniometry were the least accurate, with a minimum significant difference of 14°.
Conclusions:
- The digital inclinometer is the most accurate tool for measuring knee angles.
- The long arm goniometer is the second most accurate method.
- Visual estimation and short arm goniometry lack the accuracy required for reliable knee angle assessment.
Related Concept Videos
Reliability and Validity
13.9K
Reliability and validity are two important considerations that must be made with any type of data collection. Reliability refers to the ability to consistently produce a given result. In the context of psychological research, this would mean that any instruments or tools used to collect data do so in consistent, reproducible ways.
13.9K
Improving Translational Accuracy
14.9K
Base complementarity between the three base pairs of mRNA codon and the tRNA anticodon is not a failsafe mechanism. Inaccuracies can range from a single mismatch to no correct base pairing at all. The free energy difference between the correct and nearly correct base pairs can be as small as 3 kcal/ mol. With complementarity being the only proofreading step, the estimated error frequency would be one wrong amino acid in every 100 amino acids incorporated. However, error frequencies observed in...
14.9K
Improving Translational Accuracy
3.7K
3.7K
Uncertainty in Measurement: Accuracy and Precision
101.2K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.
101.2K
Knee Joint
3.2K
The knee joint is the most complicated joint in the body. It consists of three articulations– two tibiofemoral and one patellofemoral. As is characteristic of synovial joints, the knee joint has a thin articular capsule that partially surrounds this joint cavity. Additionally, several ligaments, muscles, and cartilaginous structures support the movement of the knee.
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
3.2K
Accuracy and Precision
15.3K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value. Highly accurate...
15.3K