Novel In Vitro Test Systems and Insights for Transcatheter Mitral Valve Design, Part I: Paravalvular Leakage
Eric L Pierce1, Vahid Sadri1, Beatrice Ncho1
1Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, 387 Technology Circle NW, Suite 200, Atlanta, GA, 30313, USA.
Abstract:
While transcatheter mitral valve (TMV) replacement technology has great clinical potential for surgically inoperable patients suffering from mitral regurgitation, no TMV has yet achieved regulatory approval. The diversity of devices currently under development reflects a lack of consensus regarding optimal design approaches. In Part I of this two-part study, a test system was developed for the quantification of paravalvular leakage (PVL) following deployment of a TMV or TMV-like device in pressurized, explanted porcine hearts (N = 7). Using this system, PVL rate was investigated as a function of steady trans-mitral pressure (ΔP), TMV shape, and TMV-annular oversizing, using a series of "mock TMV plug" devices. Across all devices, PVL was found to approximately trend with the square of ΔP. PVL rates were approximately 0-15 mL/s under hypotensive pressure, 10-40 mL/s under normotension, and 30-85 mL/s under severe hypertension. D-shaped devices significantly reduced PVL vs. circular devices; however, this effect was diminished upon oversizing to the annulus by 6 mm inter-trigonal distance. In conclusion, this steady pressure, in vitro test system was effective to compare PVL performance across TMV-like designs. PVL exhibited complex dynamics in terms of its response to transvalvular pressure and TMV profile.
More Related Videos
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Related Concept Videos
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Stenosis I: Introduction
Mitral Regurgitation I: Introduction