Kerma area product (KAP) and scatter measurements for intraoral X-ray machines using three different types of round
Dennise Magill1, Nhan James Huu Ngo2, Marc A Felice1
11 Department of Environmental Health and Radiation Safety, University of Pennsylvania , Philadelphia, PA , USA.
Objectives::
The objective of the study was to determine the radiation dose reduction achieved when rectangular collimation was used on various round collimators. In addition, we evaluated the tissue doses imparted to various head and neck organs.
Methods::
To evaluate the variation in radiation output based on the variable geometric configurations, the kerma area product (KAP) was measured using a commercially available KAP-meter with an internal ion chamber capable of detecting both radiation dose (µGy) and the primary X-ray beam area. The KAP was measured using standard 20.4, 25.7, and 31.7 cm2 round collimators with and without rectangular X-ray field restrictors. To evaluate the potential change in patient scatter radiation dose, an adult head phantom was loaded with thin strips of gafchromic film. A full mouth X-ray series was acquired with various geometric configurations. The films were quantified using a calibration factor to yield absorbed organ doses for the eyes, thyroid, and salivary glands.
Results::
With the use of rectangular collimator, the KAP for a 31.7 cm2 round collimator was reduced by up to 60% while the 20.4 cm2 round collimator elicited a reduction from up to 40%. In the organ study, results of up to 81% reduction in scatter radiation dose were observed.
Conclusions::
Although, US FDA regulations allow a maximum beam size of 38.5 cm2 on the patient skin, this study suggests that the use of rectangular collimators provide clinically relevant dose reduction for patients, even when using smaller round collimation, hence the use of rectangular collimation for all intraoral radiographic procedures is highly recommended.
More Related Videos
10:16X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
Published on: August 20, 2019
07:55Elemental-sensitive Detection of the Chemistry in Batteries through Soft X-ray Absorption Spectroscopy and Resonant Inelastic X-ray Scattering
Published on: April 17, 2018
Related Concept Videos
Distribution of Stresses in a Narrow Rectangular Beam
Types of Limits II
Types of Limits I
Limiting Reactant
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...