Fast technetium-99m liver SPECT for evaluation of the pretreatment procedure for radioembolization dosimetry
Related Experiment Video
Updated: Feb 3, 2026

Y-90 Radioembolization and PD-1 Inhibitor as Neoadjuvant Treatment in Hepatocellular Carcinoma
Published on: May 24, 2024
Sandra van der Velden1,2, Martijn M A Dietze1,2, Max A Viergever2
1Radiology and Nuclear Medicine, University Medical Center Utrecht, P.O. Box 85500, 3508 GA, Utrecht, Netherlands.
Purpose:
The efficiency of radioembolization procedures could be greatly enhanced if results of the 99m Tc-MAA pretreatment procedure were immediately available in the interventional suite, enabling 1-day procedures as a result of direct estimation of the hepatic radiation dose and lung shunt fraction. This would, however, require a relatively fast, but still quantitative, SPECT procedure, which might be achieved with acquisition protocols using nonuniform durations of the projection images.
Methods:
SPECT liver images of the 150-MBq 99m Tc-MAA pretreatment procedure were simulated for eight different lesion locations and two different lesion sizes using the digital XCAT phantom for both single- and dual-head scanning geometries with respective total acquisition times of 1, 2, 5, 10, and 30 min. Three nonuniform projection-time acquisition protocols ("half-circle SPECT (HCS)," "nonuniform SPECT (NUS) I," and "NUS II") for fast quantitative SPECT of the liver were designed and compared with the standard uniform projection-time protocol. Images were evaluated in terms of contrast-to-noise ratio (CNR), activity recovery coefficient (ARC), tumor/non-tumor (T/N) activity concentration ratio, and lung shunt fraction (LSF) estimation. In addition, image quality was verified with a physical phantom experiment, reconstructed with both clinical and Monte Carlo-based reconstruction software.
Results:
Simulations showed no substantial change in image quality and dosimetry by usage of a nonuniform projection-time acquisition protocol. Upon shortening acquisition times, CNR dropped, but ARC, T/N ratio, and LSF estimates were stable across all simulated acquisition times. Results of the physical phantom were in agreement with those of the simulations.
Conclusion:
Both uniform and nonuniform projection-time acquisition liver SPECT protocols yield accurate dosimetric metrics for radioembolization treatment planning in the interventional suite within 10 min, without compromising image quality. Consequently, fast quantitative SPECT of the liver in the interventional suite is feasible.
More Related Videos
10:54Preparation and Evaluation of 99mTc-labeled Tridentate Chelates for Pre-targeting Using Bioorthogonal Chemistry
Published on: February 4, 2017
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Fast Fourier Transform
The computational efficiency of the FFT becomes...
Fast Decoupled and DC Powerflow
Gene Evolution - Fast or Slow?
In contrast, regions which code...
Gene Evolution - Fast or Slow?
Liver Histology
Hepatocytes perform a variety of essential functions. They secrete...
Liver Physiology
Metabolic Regulation:
The liver is the central organ involved in regulating blood composition. It stabilizes blood glucose levels, maintaining them within the range of 70–110 mg/dL. When these levels drop, the liver breaks down glycogen reserves and releases glucose into the bloodstream. It can...