Related Experiment Video
Updated: Feb 3, 2026

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
High-frequency spinal cord stimulation causing cardiac paresthesias after lead migration: a case report
Neil Majmundar1, Eleonora Francesca Spinazzi1, Joseph Doran1
1Department of Neurological Surgery, Neurological Institute of New Jersey, Rutgers New Jersey Medical School, Newark, NJ, USA.
Introduction:
High-frequency (HF) spinal cord stimulation (SCS), a relatively new form of spinal cord stimulation, provides stimulation frequencies of up to 10 kHz and allows for paresthesia-free pain relief, an advantage that distinguishes it from traditional stimulation therapy. Without paresthesias, patients with HF SCS do not experience position-dependent painful stimulation and do not have to experience treatment interruption during sleep. Lead migration is a well-known complication of conventional spinal cord stimulation and usually results in a loss of efficacy along with other unpleasant sensory symptoms. In this case report, we present an incidence of lead migration in HF SCS that resulted in paresthesias, a symptom not expected to occur in this novel therapy.
Case:
The patient, a 60-year-old female with post-laminectomy syndrome, underwent a trial of HF SCS with standard lead placement at T8-T9. She initially had pain relief, but returned to the office on post-operative day 2 complaining of left chest wall and cardiac paresthesias, without frank pain or palpitations, in addition to loss of efficacy for her back and leg pain. Imaging showed that the leads had migrated, with one lead reaching the levels of T1-T3.
Conclusion:
While HF SCS has emerged as an effective paresthesia-free means of reducing back and leg pain, we provide the first report of paresthesias occurring with the HF SCS system as a result of cephalad lead migration. As HF SCS is only now being utilized as a treatment modality, we must remain cautious of potential adverse outcomes in patients, in particular above T8.
More Related Videos
08:07Paradigms of Lower Extremity Electrical Stimulation Training After Spinal Cord Injury
Published on: February 1, 2018
06:55Author Spotlight: Advancing Spinal Cord Stimulation - Exploring the Cellular Responses of Motor Neurons Through Patch-Clamp Electrophysiology
Published on: September 8, 2023
Related Concept Videos
Spinal Cord
The Spinal Cord
Spinal Cord: Information Processing
Sensory Information Processing
Sensory information processing begins at the sensory receptors located in the skin and other tissues, which detect somatic sensory stimuli such as touch, temperature, or pain. These receptors function as catalysts, initiating...
Muscle Stimulation Frequency
Wave summation
At low firing rates, motor neurons induce individual twitch contractions in muscle fibers. These twitches...
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Spinal Cord: Gross Anatomy