A multicenter study to evaluate automated platelet aggregometry on Sysmex CS-series coagulation analyzers-preliminary
Sean Platton1, Áine McCormick2, Musfira Bukht1
1The Royal London Hospital Haemophilia Centre Barts Health NHS Trust The Royal London Hospital London UK.
This study shows an automated method using Sysmex CS-2x00 analyzers can effectively measure platelet function. This automated light transmission aggregometry (LTA) offers a viable alternative to traditional, labor-intensive methods for platelet disorder detection.
Area of Science:
- Hematology
- Clinical Chemistry
- Laboratory Medicine
Background:
- Traditional light transmission aggregometry (LTA) for platelet function testing is time-consuming and labor-intensive.
- A multicenter study was conducted to evaluate an automated LTA method.
- The study aimed to establish reference ranges and an optimal testing regimen using a coagulation analyzer.
Purpose of the Study:
- To evaluate an automated light transmission aggregometry (LTA) method.
- To establish reference ranges for platelet function testing using a coagulation analyzer.
- To determine the ideal testing regimen for automated LTA.
Main Methods:
- Sysmex CS-2x00 series analyzers were utilized for automated LTA.
- A range of agonists including ADP, arachidonic acid, collagen, ristocetin, epinephrine, TRAP, and U46619 were tested at various concentrations.
- Key aggregation parameters such as maximum aggregation, final aggregation, disaggregation, slope, and acquisition time were analyzed.
Main Results:
- No significant differences in aggregation parameters were observed for several tested concentrations of ADP, arachidonic acid, collagen, ristocetin, epinephrine, and TRAP in normal subjects.
- Maximum platelet aggregation was achieved within 300 seconds for ADP, U46619, collagen, ristocetin, and TRAP at specific concentrations.
- Other agonists required 600 seconds for maximum aggregation.
Conclusions:
- A standardized panel of agonists and concentrations is suitable for automated LTA on Sysmex CS-2x00 analyzers.
- Recommended observation times are 600 seconds for most agonists and 300 seconds for TRAP and U46619.
- Automated analyzers could potentially replace dedicated aggregometers, expanding LTA availability.
More Related Videos
05:49Procoagulant Platelet Characterization by Measuring Phosphatidylserine Exposure and Microvesicle Release from Human Purified Platelets
Published on: November 29, 2024
09:13Turbidimetry on Human Washed Platelets: The Effect of the Pannexin1-inhibitor Brilliant Blue FCF on Collagen-induced Aggregation
Published on: April 6, 2017
Related Concept Videos
Coagulation
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
Finding the Center of Gravity
Resistors In Series
In a series circuit, the...
Series Resonance
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...