Comparison of intraoperative 2D vs. 3D imaging in open reduction and fixation of distal radius fractures
Diego Hammerle1, Georg Osterhoff2, Florin Allemann3
1Klinik für Orthopädische Chirurgie und Traumatologie des Bewegungsapparates, Kantonsspital St.Gallen, Rorschacher Strasse 95, 9007, St.gallen, Switzerland. Diego.hammerle@kssg.ch.
Purpose:
In the volar plating of distal radius fractures, intraoperative three-dimensional (3D) imaging is designed to allow better judgment regarding screw and implant positioning compared with conventional intraoperative two-dimensional (2D) imaging. We evaluated the impact of these two imaging modalities on the rates of intraoperative revision and secondary surgery, as well as the need for implant removal during follow-up.
Methods:
A retrospective analysis of consecutive patients who underwent volar plate osteosynthesis for isolated distal radius fractures between January 2008 and April 2016 was performed. Patient files were evaluated for intraoperative imaging findings, intraoperative and postoperative revision rates, and implant removal during follow-up. Additional analyses of radiation exposure, operation time, and hospitalization time were performed.
Results:
A total of 314 patients were analyzed (mean age: 54 ± 19 years; 210 females). For 246 patients, only 2D imaging was performed, while the remaining 68 patients underwent both 2D and 3D imaging (O-Arm, Medtronic). The intraoperative revision rate was significantly (p < 0.001) higher with 3D imaging (32.4%) compared with 2D imaging (2.0%). The postoperative revision rates were similar between both the groups (2.9% vs. 2.0%; p = 0.674). Compared with 2D imaging, the use of the Medtronic O-Arm resulted in a significantly lower implant removal rate (8.8% vs. 18.7%; p = 0.036) during follow-up.
Conclusion:
Compared with conventional 2D imaging, the use of intraoperative 3D imaging significantly increased the intraoperative revision rate and has the potential for positive long-term effects for lowering the risk of requiring an implant removal.
More Related Videos
Related Concept Videos
Radius of Gyration of an Area
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a...
Schwarzschild Radius and Event Horizon
The minimum speed required to launch a projectile from the surface of an object to which it is gravitationally bound so that it eventually escapes the object’s gravitational field is called the escape velocity. The escape velocity is independent of the mass of the object. Merging the idea of escape...
Degree of Curvature and Radius of Curvature
Oxidation-Reduction Reactions
The Sense of Self: Reflected Self-Appraisal and Social Comparison