[MRI in patients with auditory implants equipped with implanted magnets-an update : Overview and procedural
S Nospes1, M A Brockmann2, A Läßig3
1Schwerpunkt Kommunikationsstörungen/HNO-Klinik, Universitätsmedizin Mainz, Langenbeckstr. 1, 55131, Mainz, Deutschland. sabine.nospes@unimedizin-mainz.de.
Magnetic resonance imaging (MRI) is generally safe for patients with auditory implants when manufacturer guidelines are followed. However, potential risks like implant dislocation and reduced image quality necessitate careful consideration and consultation.
Area of Science:
- Otorhinolaryngology
- Medical Imaging
- Biomedical Engineering
Background:
- Over 100,000 patients in Germany use cochlear implants, with numerous others having various auditory implants containing magnets.
- Increasing use of magnetic resonance imaging (MRI) necessitates understanding implant compatibility.
- MRI compatibility is a critical safety and quality feature for auditory implant devices.
Purpose of the Study:
- To review technical innovations and new auditory devices certified in the EU and USA since November 2013.
- To assess the MRI compatibility and safety of modern auditory implants.
- To provide guidance on MRI procedures for patients with these devices.
Main Methods:
- Selective literature search in PubMed using keywords: MRI compatibility/safety + cochlear implant/auditory brainstem implant/Bonebridge/Sophono alpha/Vibrand Soundbridge/BAHA attract.
- Inclusion of all relevant publications and cited papers concerning MRI compatibility of hearing implants.
- Focus on devices with medical-technical certification in the EU and USA.
Main Results:
- High electromagnetic fields in MRI can rarely cause implant dislocation, potentially leading to device malfunction.
- Image artifacts in head MRIs are possible but can be mitigated with specific MRI sequences.
- Current auditory implants are licensed for 1.5 T MRI, with some approved up to 3.0 T.
Conclusions:
- MRI scans are safe for patients with certified auditory implants if manufacturer instructions are followed.
- Potential risks include implant defects/dislocations and reduced MRI image quality.
- MRI indication requires close consultation between physicians, implanting teams, and radiologists, after exhausting other diagnostic options.
More Related Videos
Related Concept Videos
Nuclear Magnetic Resonance (NMR): Overview
NMR spectroscopy generates a spectrum where the characteristic absorption frequencies of the sample are...
Leveling Equipment
Cardiac Catheterization I: Pre-Procedure Overview
Pharmacokinetics in Pediatric Patients: Overview and Drug Absorption
The Auditory Ossicles
The aptly named stapes look very much like a stirrup. The three ossicles are unique to mammals, and each plays a role in...
Personal Protective Equipment