Effect of Baseline Impedance on Ablation Lesion Dimensions: A Multimodality Concept Validation From Physics to
Michael Barkagan1, Markus Rottmann1, Eran Leshem1
1Cardiovascular Division, Department of Medicine, Beth Israel Deaconess Medical Center, Harvard-Thorndike Electrophysiology Institute, Harvard Medical School, Boston, MA (M.B., M.R., E.L., A.E.B., E.A.).
Background:
Radiofrequency ablation using irrigated catheters is performed using a power-controlled mode. However, lesion size is dependent on current delivery at a particular impedance, such that a power value alone may not reflect actual energy delivery, resulting in lesion size variability at similar power settings. We hypothesized that modulating baseline impedance at fixed power settings affects ablation lesion dimensions.
Methods:
In 20 ex vivo swine hearts, radiofrequency ablation was performed using an irrigated catheter at a fixed power setting of 30 W per 20 seconds and a multistepped impedance load (100-210Ω). In 4 in vivo thigh muscle preparations and right atria, ablation was performed using similar power settings at 3 baseline impedances: low (90-130Ω), intermediate (131-180Ω), and high (181-224Ω). The relationship between baseline impedance, current, and lesion dimensions was examined.
Results:
Baseline impedance had a strong negative correlation with current squared ( I2) for all experimental models: ex vivo (R=-0.94; P<0.0001), thigh muscle (R=-0.93; P<0.0001), and right atria (R=-0.94; P<0.0001). Lesion dimensions at similar power settings were highly variable and directly related to I2 (width [R=0.853], depth [R=0.814]). In the thigh muscle, lesion depth was 8.2±0.7, 6.5±0.8, and 4.2±0.5 mm for low, intermediate, and high impedance, respectively ( P<0.0001). In right atria lines, low baseline impedance resulted in wider lines (7.2±1.4 mm) relative to intermediate (5.8±1.8 mm) and high impedance (4.7±1.7 mm; P<0.0001).
Conclusions:
Radiofrequency ablation in a power control mode results in variable lesion dimensions that are partially related to differences in baseline impedance and current output. Ablation at a lower baseline impedance results in increased current output and lesion dimensions.
Related Concept Videos
Self-Concept
Infancy and Emerging Recognition
During infancy, self-concept is virtually nonexistent. Babies do not distinguish themselves as separate entities and often mistake their...
Reliability and Validity
Concepts and Prototypes
The brain organizes this information using concepts, which are mental categories grouping linguistic data,...
Support Reactions in Three Dimensions
Ball and Socket Joint is one of the supports allowing free rotation about any axis. This freedom of rotation is...
Relative Velocity in One Dimension
Formula Mass and Mole Concepts of Compounds