The Intra-Aortic Balloon Pump
Published on: February 5, 2021
Abdominal Positioning of the Next-Generation Intra-Aortic Fluid Entrainment Pump (Aortix) Improves Cardiac Output in
Shiva K Annamalai1, Michele L Esposito1, Lara A Reyelt1
1The Acute Mechanical Circulatory Support Working Group; The Molecular Cardiology Research Institute, Surgical and Interventional Research Laboratories, and the CardioVascular Center, Tufts Medical Center, Boston, MA.
Background:
Acute heart failure refractory to medical therapy is a major cause of morbidity and mortality. The Aortix device (Procyrion Inc) is a percutaneously delivered entrainment pump positioned in the descending aorta.
Methods And Results:
Using the newest generation Aortix device in 8 adult male Yorkshire swine, we tested the hypothesis that positioning in the abdominal aorta may provide superior hemodynamic effects than thoracic positioning in a swine model of postinfarct left ventricular injury.Abdominal activation generated significantly larger transaortic gradients (proximal minus distal mean aortic pressures) than thoracic positioning at all pump speeds. Compared with baseline values, activation in the abdominal, not thoracic, position significantly increased cardiac output, reduced arterial elastance, and systemic vascular resistance at low speeds. Compared with baseline values, abdominal activation also increased transpulmonary pressure gradients at medium and high speed, which was driven by trends toward higher mean pulmonary artery pressure and lower pulmonary capillary wedge pressure.
Conclusions:
This is the first report to determine that in contrast to thoracic positioning, abdominal positioning of the newest generation Aortix device reduces left ventricular afterload and increases cardiac output at low speeds. These findings have potentially important implications for the design of early clinical studies by suggesting that device position and speed are major determinants of improved hemodynamic efficacy.
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Heart Failure II: Pathophysiology