Related Experiment Video
Updated: Feb 3, 2026

Design of a Biocompatible Drug-Eluting Tracheal Stent in Mice with Laryngotracheal Stenosis
Published on: January 21, 2020
Unselected Use of Ultrathin Strut Biodegradable Polymer Sirolimus-Eluting Stent Versus Durable Polymer
Kyohei Yamaji1, Thomas Zanchin1, Christian Zanchin1
1Swiss Cardiovascular Center Bern, Department of Cardiology, Bern University Hospital, Switzerland (K.Y., T.Z., C.Z., S.S., K.C.K., L.H., F.P., S.B., C.M., A.M., C.S., M.V., M.B., T.P., S.W., L.R.).
Background:
Biodegradable polymer sirolimus-eluting stents (BP-SESs) have been reported to be noninferior compared with durable polymer everolimus-eluting stents (DP-EES) in a randomized clinical trial. We sought to compare the efficacy and safety of an ultrathin strut BP-SES with a DP-EES in an all-comers population.
Methods And Results:
Among 7640 consecutive patients who underwent percutaneous coronary intervention between March 2011 and June 2015, 4638 patients were exclusively treated with BP-SES (N=1896; 3137 lesions) or DP-EES (N=2742; 4468 lesions). After propensity score matching within strata of clinical indications, the final study population consisted of 2902 matched patients (BP-SES 2406 lesions and DP-EES 2368 lesions). The primary device-oriented composite end point (DOCE) included cardiac death, target vessel myocardial infarction, and target lesion revascularization at 1 year. BP-SES (6.9%) was noninferior to DP-EES (8.0%) with respect to device-oriented composite end point (hazard ratio [HR], 0.85; 95% CI, 0.65-1.11; P for noninferiority <0.001; P for superiority=0.24). No differences in cardiac death (BP-SES, 2.3% versus DP-EES, 3.0%; HR, 0.76; 95% CI, 0.49-1.20; P=0.25), myocardial infarction (BP-SES, 4.6% versus DP-EES, 4.6%; HR, 1.00; 95% CI, 0.71-1.40; P=0.99), or target lesion revascularization (BP-SES, 2.8% versus DP-EES, 2.5%; HR, 1.11; 95% CI, 0.71-1.74; P=0.65) were observed. The rate of periprocedural myocardial infarction was comparable between the 2 groups (2.1% versus 2.2%; HR, 0.97; 95% CI, 0.59-1.58; P=0.89). The rate of definite stent thrombosis was similarly low throughout 1 year (BP-SES, 0.8% versus DP-EES, 0.8%; HR, 1.00; 95% CI, 0.45-2.22; P=1.00).
Conclusions:
In a consecutively enrolled percutaneous coronary intervention population reflecting routine clinical practice, BP-SES was noninferior to DP-EES for device-oriented composite end point at 1 year.
Clinical Trial Registration:
URL: https://www.clinicaltrials.gov . Unique identifier: NCT02241291.
More Related Videos
Related Concept Videos
Polymers
Polymers
High-Performance Liquid Chromatography: Elution Process
Polymer Classification: Architecture
Polymer Classification: Crystallinity
Crystalline domains are the regions where polymer chains are aligned in an orderly manner and held together in proximity by intermolecular forces. For example, chains in the crystalline domains of polyethylene and nylon are bound together by van der Waals...
Polymer Classification: Stereospecificity